Holter recordings with continuous marker annotation to evaluate pacemaker function.

Bernd Nowak1, Sophie Henry, Roland Mols, Steven Maertens, Martin Coenen, Jürgen Meyer.   

Abstract

BACKGROUND: Pacemaker marker annotations facilitate the interpretation of device behavior in addition to ECG recordings. However, they are only available in conjunction with a programmer. We studied the diagnostic value of a prototype Telemetry Holter Decoder (THD), providing continuous marker annotations on a conventional Holter.
METHODS: The study included 20 patients with VDD or DDDR pacemakers. A 24-hour Holter was performed using the THD. Marker annotations are transmitted from the pacemaker to the THD, which transforms them into analog signals, which are recorded on one of the Holter channels.
RESULTS: During a total recording time of 458 hours, high quality marker annotations were retrieved for every patient. Artefacts disturbed the recordings during 184 min (0.67%). The THD provided information not discernible on the ECG: intermittent atrial undersensing during sinus rhythm (1096 times). Atrial tachycardias, not visible on the ECG, were detected in 2 patients. The activation of tachycardia response algorithms was clearly annotated in 11,516 events. A total of 8875 PVC's occurred, 57.8% of which were classified incorrectly in the event counters as conducted or fusion beats. Atrial far-field sensing or VA conduction was demonstrated 4294 times. Electromagnetic interferences, not visible on the ECG, could be seen three times.
CONCLUSION: Recording of continuous high-quality marker annotation on a conventional Holter is feasible. The THD provides important information on device behavior, even in patients assumed to have regular device function, and shows to be clearly superior to ECG interpretation alone. Such data can be used for improved programming, troubleshooting and for the validation of new algorithms.

Entities:  

Mesh:

Year:  2002        PMID: 11844288      PMCID: PMC7027633          DOI: 10.1111/j.1542-474x.2001.tb00135.x

Source DB:  PubMed          Journal:  Ann Noninvasive Electrocardiol        ISSN: 1082-720X            Impact factor:   1.468


  12 in total

1.  Fusion or confusion on Holter recording.

Authors:  L M van Gelder; F A Bracke; M I el Gamal
Journal:  Pacing Clin Electrophysiol       Date:  1991-05       Impact factor: 1.976

2.  Pacemaker generator pseudomalfunction: an artifact of Holter monitoring.

Authors:  M D Lesh; J J Langberg; J C Griffin; M M Scheinman; C L Witherell
Journal:  Pacing Clin Electrophysiol       Date:  1991-05       Impact factor: 1.976

3.  Holter, loop recorder, and event counter capabilities of implanted devices.

Authors:  J E Waktare; M Malik
Journal:  Pacing Clin Electrophysiol       Date:  1997-10       Impact factor: 1.976

Review 4.  Pacemaker event counters: possible sources of error in calculation of AV synchrony in VDD single lead systems as an example for present limitations.

Authors:  C W Israel; J B Böckenförde
Journal:  Pacing Clin Electrophysiol       Date:  1998-03       Impact factor: 1.976

5.  Diagnosis of atrial undersensing in dual chamber pacemakers: impact of autodiagnostic features.

Authors:  U K Wiegand; F Bode; R Schneider; A Brandes; H Haase; H A Katus; J Potratz
Journal:  Pacing Clin Electrophysiol       Date:  1999-06       Impact factor: 1.976

6.  Apparent failure to sense (undersensing) caused by oversensing: diagnostic use of a noninvasively obtained intracardiac electrogram.

Authors:  M A Warnowicz; N Goldschlager
Journal:  Pacing Clin Electrophysiol       Date:  1983-11       Impact factor: 1.976

Review 7.  Taking advantage of sophisticated pacemaker diagnostics.

Authors:  B Nowak
Journal:  Am J Cardiol       Date:  1999-03-11       Impact factor: 2.778

8.  [Incidence of atrial arrhythmia in patients with long term dual-chamber pacemakers. Contribution of the Holter function of pacemakers].

Authors:  S Garrigue; S Cazeau; P Ritter; A Lazarus; D Gras; J Mugica
Journal:  Arch Mal Coeur Vaiss       Date:  1996-07

9.  Holter recordings with continuous marker annotations: a new tool in pacemaker diagnostics.

Authors:  B Nowak; T Middeldorf; C M Housworth; A Brüls; A Liebrich; S Rosocha; T Voigtländer; E Himmrich; J Meyer
Journal:  Pacing Clin Electrophysiol       Date:  1996-11       Impact factor: 1.976

10.  Diagnosis of ventricular tachycardia using a pacemaker Holter function.

Authors:  G Lascault; R Frank; C Himbert; G Fontaine; D Thomas; Y Grosgogeat
Journal:  Pacing Clin Electrophysiol       Date:  1994-07       Impact factor: 1.976

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.