Literature DB >> 11838808

Electromagnetic field hazards involving adjustable shunt valves in hydrocephalus.

Thomas Schneider1, Uwe Knauff, Jürgen Nitsch, Raimund Firsching.   

Abstract

OBJECT: Standard therapy for hydrocephalus involves shunts and valves, which are frequently adjustable. Because of increasing "electromagnetic smog" (for example, that generated by cellular phones), these valves are often exposed to electromagnetic fields.
METHODS: Various magnetic fields were tested for their effects on two different kinds of adjustable valves. The minimum magnetic flux density affecting the adjustment of the valve was determined. Results were compared with magnetic fields found in contemporary everyday life. In homogeneous magnetic fields the adjustment of one valve (Sophysa model SM8) was changed at 5 mT, whereas the second valve (Codman Hakim model CM) was not affected. In nonhomogeneous fields the SM8 valve was affected at 25 mT and the CM valve at 15 mT. Thus, these valves may be affected by headphones and telephone receivers. Surroundings such as the Japanese magnetic suspension railway and the lead cabin of electrical railway engines, in which critical levels of magnetic flux may be present, may also affect adjustable valves. The high-frequency fields of cellular phones, however, have no effect on these valves.
CONCLUSIONS: Every surgeon who implants these valves and every patient who receives them should know the possible hazards. The valve selection should be adapted to the environment of the patient. Devices with critical levels of electromagnetic flux that are used in the homes of patients should be replaced by ones with lower magnetic fields. The future construction of these valves should be modified in such a way that their adjustment requires a higher magnetic flux density, so that the valves become less sensitive to unwanted effects from environmental magnetic fields.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11838808     DOI: 10.3171/jns.2002.96.2.0331

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  11 in total

1.  The Strata programmable valve for shunt-dependent hydrocephalus: the pediatric experience at a single institution.

Authors:  Edward S Ahn; Markus Bookland; Benjamin S Carson; Jon D Weingart; George I Jallo
Journal:  Childs Nerv Syst       Date:  2006-10-07       Impact factor: 1.475

2.  Programmable CSF shunt valve: in vitro assessment of MR imaging safety at 3T.

Authors:  F G Shellock; R Habibi; J Knebel
Journal:  AJNR Am J Neuroradiol       Date:  2006-03       Impact factor: 3.825

3.  Magnetic toys: forbidden for pediatric patients with certain programmable shunt valves?

Authors:  Tycho J Zuzak; Bettina Balmer; Daniel Schmidig; Eugen Boltshauser; Michael A Grotzer
Journal:  Childs Nerv Syst       Date:  2008-12-05       Impact factor: 1.475

4.  Radiation risk due to shunted hydrocephalus and the role of MR imaging-safe programmable valves.

Authors:  S Krishnamurthy; B Schmidt; M D Tichenor
Journal:  AJNR Am J Neuroradiol       Date:  2012-11-01       Impact factor: 3.825

5.  Evaluation of magnetic resonance imaging issues for implantable microfabricated magnetic actuators.

Authors:  Hyowon Lee; Qing Xu; Frank G Shellock; Marvin Bergsneider; Jack W Judy
Journal:  Biomed Microdevices       Date:  2014-02       Impact factor: 2.838

6.  Interactions between programmable shunt valves and the iPad 3 with Smart Cover.

Authors:  Yizheng He; Rory K J Murphy; Jarod L Roland; David D Limbrick
Journal:  Childs Nerv Syst       Date:  2013-02-20       Impact factor: 1.475

7.  Management of neonatal hydrocephalus: feasibility of use and safety of two programmable (Sophy and Polaris) valves.

Authors:  Juan F Martínez-Lage; María-José Almagro; Isabel Sanchez Del Rincón; Miguel A Pérez-Espejo; Claudio Piqueras; Raúl Alfaro; Javier Ros de San Pedro
Journal:  Childs Nerv Syst       Date:  2007-10-09       Impact factor: 1.475

8.  In vitro hydrodynamic properties of the Miethke ProGAV hydrocephalus shunt.

Authors:  David M Allin; Zofia H Czosnyka; Marek Czosnyka; Hugh K Richards; John D Pickard
Journal:  Cerebrospinal Fluid Res       Date:  2006-06-29

9.  Stability of Programmable Shunt Valve Settings with Simultaneous Use of the Optune Transducer Array: A Case Report.

Authors:  Andrew K Chan; Harjus S Birk; Ethan A Winkler; Jennifer A Viner; Jennie W Taylor; Michael W McDermott
Journal:  Cureus       Date:  2016-07-07

10.  Investigation of the hydrodynamic properties of a new MRI-resistant programmable hydrocephalus shunt.

Authors:  David M Allin; Marek Czosnyka; Hugh K Richards; John D Pickard; Zofia H Czosnyka
Journal:  Cerebrospinal Fluid Res       Date:  2008-04-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.