An efficient, accurate and robust radiometer configuration for microwave temperature measurement for industrial and medical applications.

D V Land1.   

Abstract

The theoretical and practical design of a type of microwave radiometer particularly suited to industrial and medical temperature measurement is described. The use of sequential switching of source and reference signals with synchronous demodulation of switched signal components gives an efficient and robust instrument configuration. The radiometer measures the true, matched-impedance effective temperature of the coupled source. Changes in noise-figure, gain or frequency response of the radiometer microwave amplification do not affect the temperature measurement. Practical microwave radiometers using this configuration can be simply calibrated, and have inherently stable calibration and excellent accuracy.

Mesh:

Year:  2001        PMID: 11837048     DOI: 10.1080/08327823.2001.11688456

Source DB:  PubMed          Journal:  J Microw Power Electromagn Energy        ISSN: 0832-7823            Impact factor:   1.325


  2 in total

1.  Modeling the detectability of vesicoureteral reflux using microwave radiometry.

Authors:  Kavitha Arunachalam; Paolo F Maccarini; Valeria De Luca; Fernando Bardati; Brent W Snow; Paul R Stauffer
Journal:  Phys Med Biol       Date:  2010-08-25       Impact factor: 3.609

2.  Detection of vesicoureteral reflux using microwave radiometry-system characterization with tissue phantoms.

Authors:  Kavitha Arunachalam; Paolo Maccarini; Valeria De Luca; Piero Tognolatti; Fernando Bardati; Brent Snow; Paul Stauffer
Journal:  IEEE Trans Biomed Eng       Date:  2011-01-20       Impact factor: 4.538

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.