| Literature DB >> 11830375 |
J Bellemans1, H Vandenneucker, L Labey, R Van Audekercke.
Abstract
The aim of this study was to measure and compare the ultimate failure strengths and cyclic fatigue strengths of currently available meniscal suturing devices. No statistically significant difference in failure load was found between a vertical loop suture (mean 46.3 N), a horizontal mattress suture (52.5 N), the T-fix Device (47.5 N), and the 16-mm (39.2 N) and 13-mm (32.8 N) Bionix Arrow. Statistically inferior results were seen with the 10-mm Bionix Arrow (18.8 N), the S.D. Sorb Stapler (4.3 N), and the 12-mm Arthrex Meniscal Dart (10.5 N) (P<0.01). The Mitek Meniscal Repair System (28.1 N) performed intermediate, with significantly better results than the S.D. Sorb Stapler and the 12-mm Arthrex Dart, but significantly worse than the vertical and horizontal sutures, the T-fix and Bionix 16-mm Device (P<0.01). Cyclic fatigue strength was significantly less for the 10-mm Bionix Arrow, the S.D. Sorb Stapler, and the 12-mm Arthrex Meniscal Dart (P<0.01) compared to all other devices. We conclude that the 13- and 16-mm Bionix Arrow and the T-fix Device have comparable ultimate failure strengths and cyclic fatigue strengths to conventional meniscal suturing techniques. The 10-mm Bionix Arrow, S.D. Sorb Stapler, and the 12-mm Arthrex Meniscal Dart, however, have far inferior failure and cyclic fatigue strengths and their clinical application should be questioned.Entities:
Mesh:
Year: 2002 PMID: 11830375 DOI: 10.1016/s0968-0160(01)00126-0
Source DB: PubMed Journal: Knee ISSN: 0968-0160 Impact factor: 2.199