Literature DB >> 11830375

Fixation strength of meniscal repair devices.

J Bellemans1, H Vandenneucker, L Labey, R Van Audekercke.   

Abstract

The aim of this study was to measure and compare the ultimate failure strengths and cyclic fatigue strengths of currently available meniscal suturing devices. No statistically significant difference in failure load was found between a vertical loop suture (mean 46.3 N), a horizontal mattress suture (52.5 N), the T-fix Device (47.5 N), and the 16-mm (39.2 N) and 13-mm (32.8 N) Bionix Arrow. Statistically inferior results were seen with the 10-mm Bionix Arrow (18.8 N), the S.D. Sorb Stapler (4.3 N), and the 12-mm Arthrex Meniscal Dart (10.5 N) (P<0.01). The Mitek Meniscal Repair System (28.1 N) performed intermediate, with significantly better results than the S.D. Sorb Stapler and the 12-mm Arthrex Dart, but significantly worse than the vertical and horizontal sutures, the T-fix and Bionix 16-mm Device (P<0.01). Cyclic fatigue strength was significantly less for the 10-mm Bionix Arrow, the S.D. Sorb Stapler, and the 12-mm Arthrex Meniscal Dart (P<0.01) compared to all other devices. We conclude that the 13- and 16-mm Bionix Arrow and the T-fix Device have comparable ultimate failure strengths and cyclic fatigue strengths to conventional meniscal suturing techniques. The 10-mm Bionix Arrow, S.D. Sorb Stapler, and the 12-mm Arthrex Meniscal Dart, however, have far inferior failure and cyclic fatigue strengths and their clinical application should be questioned.

Entities:  

Mesh:

Year:  2002        PMID: 11830375     DOI: 10.1016/s0968-0160(01)00126-0

Source DB:  PubMed          Journal:  Knee        ISSN: 0968-0160            Impact factor:   2.199


  5 in total

1.  A new rigid biodegradable anchor for meniscus refixation: biomechanical evaluation.

Authors:  Thore Zantop; Anne Kathleen Eggers; Volker Musahl; Andre Weimann; Joachim Hassenpflug; Wolf Petersen
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-12-19       Impact factor: 4.342

2.  Histoacryl glue in meniscal repairs (a biomechanical study).

Authors:  Irfan Ayan; Mehmet Colak; Ulku Comelekoglu; Abtullah Milcan; Oya Ogenler; Volkan Oztuna; Fehmi Kuyurtar
Journal:  Int Orthop       Date:  2006-06-08       Impact factor: 3.075

3.  Biomechanical evaluation of arthroscopic all-inside meniscus repairs.

Authors:  Taner Gunes; Bora Bostan; Mehmet Erdem; Murat Asci; Cengiz Sen; Mehmet Halidun Kelestemur
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2009-07-24       Impact factor: 4.342

4.  Bonding and fusion of meniscus fibrocartilage using a novel chondroitin sulfate bone marrow tissue adhesive.

Authors:  Jacob A Simson; Iossif A Strehin; Brian W Allen; Jennifer H Elisseeff
Journal:  Tissue Eng Part A       Date:  2013-06-12       Impact factor: 3.845

5.  Complications in medial meniscus suture: a cadaveric study.

Authors:  Alejandro Espejo-Baena; Pau Golano; Stephan Meschian; Jose Maria Garcia-Herrera; J M Serrano Jiménez
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-06-07       Impact factor: 4.114

  5 in total