Improving the effectiveness of the infant contrast sensitivity card procedure.

James R Drover1, Avery E Earle, Mary L Courage, Russell J Adams.   

Abstract

Based on results from an earlier prototype, custom software and printing techniques were developed to construct a new card-based test of contrast sensitivity (CS) for nonverbal subjects. Compared with the prototype, the new CS card test contains three improvements: (1) larger, more salient test gratings; (2) higher contrast warm-up cards for each spatial frequency set; and (3) smaller contrast step size between adjacent cards. The success of the new cards was evaluated by testing 3.5- and 12-month-old human infants. Results indicated that the new version of the test required little time to complete (mean, 6.5 min) and provided accurate estimates of visual acuity. Also, group contrast sensitivity functions (CSFs) showed substantial development from 3.5 to 12 months of age. Surprisingly, however, group CSFs obtained with the new cards were lower than those obtained with the prototype, a discrepancy that may be due to differences in space average luminance between the two sets of cards. In all, the new CS card procedure possesses several merits that give it potential as an option for assessing spatial vision in infants, toddlers, and subjects with multiple impairments.

Entities:  

Mesh:

Year:  2002        PMID: 11828899     DOI: 10.1097/00006324-200201000-00013

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  4 in total

1.  Preoperative visual acuity and contrast sensitivity in children with small, partial, or non-central cataracts.

Authors:  Eileen E Birch; Vidhya Subramanian; Christina Cheng Patel; David Stager
Journal:  J AAPOS       Date:  2013-08       Impact factor: 1.220

2.  The contrast sensitivity of the newborn human infant.

Authors:  Angela M Brown; Delwin T Lindsey; Joanna G Cammenga; Peter J Giannone; Michael R Stenger
Journal:  Invest Ophthalmol Vis Sci       Date:  2015-01-06       Impact factor: 4.799

Review 3.  Contrast insensitivity: the critical immaturity in infant visual performance.

Authors:  Angela M Brown; Delwin T Lindsey
Journal:  Optom Vis Sci       Date:  2009-06       Impact factor: 1.973

4.  A New Contrast Sensitivity Test for Pediatric Patients: Feasibility and Inter-Examiner Reliability in Ocular Disorders and Cerebral Visual Impairment.

Authors:  D Luisa Mayer; Christopher Patrick Taylor; Barry S Kran
Journal:  Transl Vis Sci Technol       Date:  2020-08-20       Impact factor: 3.283

  4 in total