Effects of clothing on accuracy and reliability of air displacement plethysmography.

Jason D Vescovi1, Shannon L Zimmerman, Wayne C Miller, Bo Fernhall.   

Abstract

PURPOSE: The purpose of this investigation was to quantify the effects of different attire on the accuracy and reliability of estimating percent body fat (%fat) by using air displacement plethysmography (ADP).
METHODS: Fifteen adults were tested four times on four separate occasions. Each session consisted of estimating %fat with ADP wearing a swimsuit (ADP(SS)), a hospital gown (ADP(HG)), and in the nude (ADP(N)), plus assessment by hydrostatic weighing (HW).
RESULTS: An ANOVA revealed no significant differences within any of the four methods over the 4 days. Intraclass correlation coefficients revealed a strong relationship for repeated measures in ADP(SS) (r = 0.981), ADP(HG) (r = 0.993), ADP(N) (r = 0.989), and HW (r = 0.976). Mean data for each condition indicated a significant underestimation (P < 0.05) of %fat while wearing a hospital gown (13.8 +/- 7.7%) compared with ADP(SS) (22.0 +/- 7.8%), ADP(N) (23.5 +/- 7.5%), and HW (22.6 +/- 6.8%).
CONCLUSION: Reliability of ADP does not appear to be compromised by clothing; however, wearing a hospital gown significantly affects accuracy, reducing %fat estimations by approximately 9% compared with the recommended swimsuit. Measurement in the nude does not provide more reliable or accurate measures than wearing a swimsuit.

Entities:  

Mesh:

Year:  2002        PMID: 11828238     DOI: 10.1097/00005768-200202000-00016

Source DB:  PubMed          Journal:  Med Sci Sports Exerc        ISSN: 0195-9131            Impact factor:   5.411


  5 in total

1.  Comparative assessment of change in fat mass using dual X-ray absorptiometry and air-displacement plethysmography.

Authors:  Joshua Warolin; Jeff Kantor; Lauren E Whitaker; Leena Choi; Sari Acra; Maciej S Buchowski
Journal:  Clin Obes       Date:  2012-06-01

2.  Subject positioning in the BOD POD® only marginally affects measurement of body volume and estimation of percent body fat in young adult men.

Authors:  Maarten W Peeters
Journal:  PLoS One       Date:  2012-03-26       Impact factor: 3.240

3.  Effect of short schemes on body composition measurements using air-displacement plethysmography.

Authors:  Holly R Hull; David A Fields
Journal:  Dyn Med       Date:  2005-07-26

4.  Comparison of air displacement plethysmography to hydrostatic weighing for estimating total body density in children.

Authors:  Geo Claros; Holly R Hull; David A Fields
Journal:  BMC Pediatr       Date:  2005-09-09       Impact factor: 2.125

5.  The impact of subject positioning on body composition assessments by air displacement plethysmography evaluated in a heterogeneous sample.

Authors:  Raluca Horhat; Monica Miclos-Balica; Paul Muntean; Sandra Popa; Irina Sima; Bogdan Glisici; Onisim Cîrja; Adrian Neagu; Monica Neagu
Journal:  PLoS One       Date:  2022-04-15       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.