Literature DB >> 11764439

A retrieval study of capital hip prostheses with titanium alloy femoral stems.

L R Mcgrath1, D L Shardlow, E Ingham, M Andrews, J Ivory, M H Stone, J Fisher.   

Abstract

We have examined 26 retrieved, failed titanium-alloy femoral stems. The clinical details, radiological appearances and the histology of the surrounding soft tissues in each patient were also investigated. The stems were predominantly of the flanged design and had a characteristic pattern of wear. A review of the radiographs showed a series of changes, progressive with time. The first was lateral debonding with subsidence of the stem. This was followed by calcar resorption and fragmentation or fracture of the cement. Finally, osteolysis was seen, starting with a radiolucency at the cement-bone interface and progressing to endosteal cavitation. Three histological appearances were noted: granulomatous, necrobiotic and necrotic. We suggest that an unknown factor, possibly related to the design of the stem, caused it to move early. After this, micromovement at the cement-stem interface led to the generation of particulate debris and fracture of the cement. A soft-tissue reaction to the debris resulted in osteolysis and failure of fixation of the prostheses.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11764439     DOI: 10.1302/0301-620x.83b8.9874

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  3 in total

1.  Coordinating retrieval and register studies improves postmarket surveillance.

Authors:  Peter Ellison; Geir Hallan; Paul Johan Høl; Nils Roar Gjerdet; Leif I Havelin
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

2.  Clinical and biological assessment of cemented titanium femoral stems: an 11-year experience.

Authors:  Patrick Boyer; Jean-Yves Lazennec; Joel Poupon; Marc-Antoine Rousseau; Philippe Ravaud; Yves Catonné
Journal:  Int Orthop       Date:  2008-10-28       Impact factor: 3.075

3.  Cytokine Profile in Patients with Aseptic Loosening of Total Hip Replacements and Its Relation to Metal Release and Metal Allergy.

Authors:  Rune J Christiansen; Henrik J Münch; Charlotte M Bonefeld; Jacob P Thyssen; Jens J Sloth; Carsten Geisler; Kjeld Søballe; Morten S Jellesen; Stig S Jakobsen
Journal:  J Clin Med       Date:  2019-08-20       Impact factor: 4.241

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.