Energy and dose rate dependence of BANG-2 polymer-gel dosimeter.

J Novotny1, V Spevacek, P Dvorak, J Novotny1, T Cechak.   

Abstract

The purpose of this study was to evaluate dependence of BANG-2 polymer-gel dosimeter sensitivity on different photon and electron energies as well as on different mean dose rates expressed as repetition rates for a standard clinically used linear accelerator. The sensitivity of the dosimeter was represented by the slope of calibration curve in the linear region measured for each modality. A calibration curve (in the linear region) based on five dosimeters (four irradiated and one background) was obtained for each photon and electron energy and different repetition rates. Dosimeter sensitivity dependence on energy was studied for 4, 6, and 18 MV x-ray photons and for nominal electron energies 9, 12, 16, and 20 MeV. Dosimeter sensitivity dependence on mean dose rate was separately studied for electron and photon beams with the use of repetition rates 80, 160, 240, 320, and 400 MU min(-1). Evaluation of dosimeters was performed on Siemens MAGNETOM EXPERT 1T scanner in the head coil. A multiecho sequence with 16 equidistant echoes was used for the evaluation of irradiated polymer-gel dosimeters. The parameters of the sequence were as follows: TR 2000 ms, TE 22.5-360.0 ms, slice thickness 5 mm, FOV 255 mm, one acquisition. There was observed a trend in polymer-gel dosimeter sensitivity dependence on the quality index of high energy x-ray beams used and on mean electron energy absorbed in the center of the dosimeter. Polymer-gel dosimeter sensitivity was decreasing with increasing photon or electron energy. There was observed no trend in polymer-gel dosimeter sensitivity dependence on mean dose rate expressed as a repetition rate for both photon and electron beams.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11764046     DOI: 10.1118/1.1414307

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Sensitivity calibration procedures in optical-CT scanning of BANG 3 polymer gel dosimeters.

Authors:  Y Xu; Cheng-Shie Wuu; Marek J Maryanski
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

Review 2.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

3.  Determine the Dose Distribution Using Ultrasound Parameters in MAGIC-f Polymer Gels.

Authors:  Hossein Masoumi; Manijhe Mokhtari-Dizaji; Azim Arbabi; Mohsen Bakhshandeh
Journal:  Dose Response       Date:  2016-02-11       Impact factor: 2.658

Review 4.  Radiation Dosimetry by Use of Radiosensitive Hydrogels and Polymers: Mechanisms, State-of-the-Art and Perspective from 3D to 4D.

Authors:  Yves De Deene
Journal:  Gels       Date:  2022-09-19
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.