1H spectroscopy without solvent suppression: characterization of signal modulations at short echo times.

D B Clayton1, M A Elliott, J S Leigh, R E Lenkinski.   

Abstract

While most proton ((1)H) spectra acquired in vivo utilize selective suppression of the solvent signal for more sensitive detection of signals from the dilute metabolites, recent reports have demonstrated the feasibility and advantages of collecting in vivo data without solvent attenuation. When these acquisitions are performed at short echo times, the presence of frequency modulations of the water resonance may become an obstacle to the identification and quantitation of metabolite resonances. The present report addresses the characteristics, origin, and elimination of these sidebands. Sideband amplitudes were measured as a function of delay time between gradient pulse and data collection, as a function of gradient pulse amplitude, and as a function of spatial location of the sample for each of the three orthogonal gradient sets. Acoustic acquisitions were performed to demonstrate the correlation between mechanical vibration resonances and the frequencies of MR sidebands. A mathematical framework is developed and compared with the experimental results. This derivation is based on the theory that these frequency modulations are induced by magnetic field fluctuations generated by the transient oscillations of gradient coils.

Entities:  

Mesh:

Year:  2001        PMID: 11740895     DOI: 10.1006/jmre.2001.2442

Source DB:  PubMed          Journal:  J Magn Reson        ISSN: 1090-7807            Impact factor:   2.229


  12 in total

1.  The rapid and automatic combination of proton MRSI data using multi-channel coils without water suppression.

Authors:  Zhengchao Dong; Bradley Peterson
Journal:  Magn Reson Imaging       Date:  2007-02-20       Impact factor: 2.546

Review 2.  Metabolite quantification and high-field MRS in breast cancer.

Authors:  Ihab S Haddadin; Adeka McIntosh; Sina Meisamy; Curt Corum; Angela L Styczynski Snyder; Nathaniel J Powell; Michael T Nelson; Douglas Yee; Michael Garwood; Patrick J Bolan
Journal:  NMR Biomed       Date:  2009-01       Impact factor: 4.044

3.  Compensation of gradient-induced magnetic field perturbations.

Authors:  Terence W Nixon; Scott McIntyre; Douglas L Rothman; Robin A de Graaf
Journal:  J Magn Reson       Date:  2008-02-23       Impact factor: 2.229

4.  Proton CSI without solvent suppression with strongly reduced field gradient related sideband artifacts.

Authors:  Grzegorz L Chadzynski; Uwe Klose
Journal:  MAGMA       Date:  2012-08-29       Impact factor: 2.310

5.  On the accurate analysis of vibroacoustics in head insert gradient coils.

Authors:  Simone A Winkler; Andrew Alejski; Trevor Wade; Charles A McKenzie; Brian K Rutt
Journal:  Magn Reson Med       Date:  2016-11-17       Impact factor: 4.668

6.  Multi-spin-echo J-resolved spectroscopic imaging without water suppression: application to a rat glioma at 7 T.

Authors:  G Hérigault; S Zoula; C Rémy; M Décorps; A Ziegler
Journal:  MAGMA       Date:  2004-10-26       Impact factor: 2.310

7.  Absolute choline tissue concentration mapping for prostate cancer localization and characterization using 3D 1 H MRSI without water-signal suppression.

Authors:  Nassim Tayari; Alan J Wright; Arend Heerschap
Journal:  Magn Reson Med       Date:  2021-09-23       Impact factor: 3.737

8.  High dynamic-range magnetic resonance spectroscopy (MRS) time-domain signal analysis.

Authors:  William C Hutton; G Larry Bretthorst; Joel R Garbow; Joseph J H Ackerman
Journal:  Magn Reson Med       Date:  2009-10       Impact factor: 4.668

9.  Field camera versus phantom-based measurement of the gradient system transfer function (GSTF) with dwell time compensation.

Authors:  M Stich; J A J Richter; T Wech; T A Bley; R Ringler; H Köstler; A E Campbell-Washburn
Journal:  Magn Reson Imaging       Date:  2020-06-10       Impact factor: 2.546

Review 10.  Motion correction in magnetic resonance spectroscopy.

Authors:  Muhammad G Saleh; Richard A E Edden; Linda Chang; Thomas Ernst
Journal:  Magn Reson Med       Date:  2020-04-17       Impact factor: 3.737

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.