O C Maguire1, M P Curry, P O'Gorman, N Parfrey, J Hegarty, S K Cunningham.
Abstract
Total complement 4 (C4) levels, when analysed on the Beckman Array nephelometer, were found to increase number of serum specimens [predominantly from patients with hepatitis C virus (HCV) infection] after overnight storage at 4 degrees C. In order to investigate whether the phenomenon of in vitro cold-dependent activation of complement (CDAC) was the explanation for this increase, paired specimens were collected from 63 patients with HCV infection in tubes with no anticoagulant (serum) and in tubes containing EDTA (which inhibits complement activation). C4 levels increased after overnight storage at 4 degrees C in 33 serum specimens (52%). In contrast, no increase in C4 levels was observed in any of the 63 EDTA specimens. Immunofixation of intact and activated C4 products confirmed that complement activation had taken place in the serum specimens in which C4 levels had increased after storage. There was a higher frequency of hepatitis C viraemia (P<0.0001), HCV antibody positivity (P<0.05) and the presence of rheumatoid factor (P<0.05) in the group of patients in whose serum samples CDAC had occurred (n = 33) than in the other group (n = 30). As a result of our findings on C4 analysis in stored serum specimens, we would recommend potassium EDTA plasma as the specimen of choice for complement analysis on the Beckman Array.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11732652 DOI: 10.1258/0004563011900902