Effect of ambient temperature on analytical and clinical performance of a blood glucose monitoring system: Omnitest Sensor glucose meter.

H Nawawi1, B S Sazali, B H Kamaruzaman, T N Yazid, A A Jemain, F Ismail, B A Khalid.   

Abstract

The effect of ambient temperature on the analytical and clinical performance of a glucose meter was examined. A total of 114 venous whole blood samples were analysed for glucose by a reference method, and by a glucose meter at 21-22 degrees C, room temperatures, 26-27 degrees C and 33-34 degrees C. Glucose meter readings at each temperature were compared with the reference values and evaluated by analysis of variance, Spearman's correlation, the percentage of glucose meter readings within +/- 10% of the reference value and error grid analysis. Analysis of covariance was used to determine the effect of temperature on glucose meter readings. There were no significant differences in the glucose meter readings and in accuracy of the meter readings between different temperatures. Temperature was not a significant independent determinant of the glucose meter readings. For each glucose concentration, the precision of the meter and clinical performance were comparable between the different temperatures. In conclusion, ambient temperature does not affect the accuracy, precision and clinical performance of the Omnitest Sensor.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11732650     DOI: 10.1258/0004563011901091

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  3 in total

1.  Hypoglycemia at admission is associated with inhospital mortality in Ugandan patients with severe sepsis.

Authors:  Richard Ssekitoleko; Shevin T Jacob; Patrick Banura; Relana Pinkerton; David B Meya; Steven J Reynolds; Nathan Kenya-Mugisha; Harriet Mayanja-Kizza; Rose Muhindo; Sanjay Bhagani; W Michael Scheld; Christopher C Moore
Journal:  Crit Care Med       Date:  2011-10       Impact factor: 7.598

Review 2.  Challenges to glycemic measurement in the perioperative and critically ill patient: a review.

Authors:  Andrew D Pitkin; Mark J Rice
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

3.  The accuracy of self monitoring blood glucose meter systems in Kampala Uganda.

Authors:  Gabriel S Bimenya; G R Nzarubara; J Kiconco; S Sabuni; W Byarugaba
Journal:  Afr Health Sci       Date:  2003-04       Impact factor: 0.927

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.