Literature DB >> 11723078

A pilot study of the continuous glucose monitoring system: clinical decisions and glycemic control after its use in pediatric type 1 diabetic subjects.

F R Kaufman1, L C Gibson, M Halvorson, S Carpenter, L K Fisher, P Pitukcheewanont.   

Abstract

OBJECTIVE: To determine whether the continuous glucose monitoring system (CGMS) (MiniMed, Sylmar, CA) could be used to make clinical decisions and whether it has an impact on glycemia in pediatric type 1 diabetic subjects. RESEARCH DESIGN AND METHODS: Pediatric subjects were recruited if they had HbA(1c) >8.0% with management problems (n = 35) or episodes of severe or nocturnal hypoglycemia or hypoglycemia unawareness associated with HbA(1c) < or =8.0% (n = 12). A total of 47 patients with a mean HbA(1c) value of 8.6 +/- 1.6% (mean age 11.8 +/- 4.6 years, youngest 2.7 years, and diabetes duration 5.5 +/- 3.5 years) on three to four insulin injections/day (n = 24) or insulin pump therapy (n = 23) were followed with the CGMS for a mean of 69.5 +/- 28 h. Comparisons were made between the number of high (>150 mg/dl) and low (<70 mg/dl) glucose patterns discerned with the sensor or the logbook, and HbA(1c) levels were evaluated.
RESULTS: In patients on injection therapy, 30 high or low glucose patterns were discerned with the logbook records and 120 patterns with the CGMS. Specific alterations of the diabetes regimen were made. An overall significant change in HbA(1c), from 3 months before wearing the sensor to 6 months after (analysis of variance 0.04), was found in the subjects. Post hoc analysis showed a significant change in HbA(1c) from 8.6 +/- 1.5% at baseline to 8.4 +/- 1.3% at 3 months (paired Student's t test 0.03).
CONCLUSIONS: The CGMS can be used by pediatric patients to detect abnormal patterns of glycemia. The information that was obtained could be used to alter the diabetes regimen and impact glycemic outcome.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11723078     DOI: 10.2337/diacare.24.12.2030

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  40 in total

1.  Accuracy of the Sof-sensor glucose sensor with the iPro calibration algorithm.

Authors:  John B Welsh; Francine R Kaufman; Scott W Lee
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 2.  Continuous glucose monitoring-guided insulin dosing in pump-treated patients with type 1 diabetes: a clinical guide.

Authors:  Allen B King
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

3.  Professional continuous glucose monitoring in clinical practice 2010.

Authors:  Thomas C Blevins
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

4.  The Contemporary Role of Masked Continuous Glucose Monitoring in a Real-Time World.

Authors:  Ian Blumer
Journal:  J Diabetes Sci Technol       Date:  2016-05-03

5.  In vivo, transcutaneous glucose sensing using surface-enhanced spatially offset Raman spectroscopy: multiple rats, improved hypoglycemic accuracy, low incident power, and continuous monitoring for greater than 17 days.

Authors:  Ke Ma; Jonathan M Yuen; Nilam C Shah; Joseph T Walsh; Matthew R Glucksberg; Richard P Van Duyne
Journal:  Anal Chem       Date:  2011-11-02       Impact factor: 6.986

6.  Use of continuous glucose monitoring in young children with type 1 diabetes: implications for behavioral research.

Authors:  Susana R Patton; Laura B Williams; Sally J Eder; Megan J Crawford; Lawrence Dolan; Scott W Powers
Journal:  Pediatr Diabetes       Date:  2011-02       Impact factor: 4.866

7.  Effects of pioglitazone and insulin on tight glycaemic control assessed by the continuous glucose monitoring system : a monocentric, parallel-cohort study.

Authors:  Wolfgang Jung; Sigrun Jung
Journal:  Clin Drug Investig       Date:  2005       Impact factor: 2.859

8.  Continuous glucose monitoring in non-insulin-using individuals with type 2 diabetes: acceptability, feasibility, and teaching opportunities.

Authors:  Nancy A Allen; James A Fain; Barry Braun; Stuart R Chipkin
Journal:  Diabetes Technol Ther       Date:  2009-03       Impact factor: 6.118

9.  Real-time continuous glucose monitoring in the clinical setting: the good, the bad, and the practical.

Authors:  Irene Mamkin; Svetlana Ten; Sonal Bhandari; Neesha Ramchandani
Journal:  J Diabetes Sci Technol       Date:  2008-09

10.  Professional continuous glucose monitoring in subjects with type 1 diabetes: retrospective hypoglycemia detection.

Authors:  Morten Hasselstrøm Jensen; Toke Folke Christensen; Lise Tarnow; Zeinab Mahmoudi; Mette Dencker Johansen; Ole Kristian Hejlesen
Journal:  J Diabetes Sci Technol       Date:  2013-01-01
View more