J V Trapp1, S A Bäck, M Lepage, G Michael, C Baldock.
Abstract
Changes in the linear attenuation coefficient of polymer gel dosimeters post-irradiation enable the imaging of dose distributions by x-ray computed tomography (CT). Various compositions of polymer gel dosimeters manufactured from acrylamide (AA), and N,N'-methylene-bis-acrylamide (BIS) comonomers and gelatin or agarose gelling agents were investigated. This work shows that increasing the comonomer concentration increases the CT-dose sensitivity of the polymer gel dosimeter. This can be further increased by replacing gelatin with agarose. Varying the gelatin concentration however does not significantly change the CT-dose sensitivity. Among the compositions studied, dose resolution (D(delta)95%) was found to be optimal for polymer gel dosimeters comprising 5% gelatin, 3% AA, 3% BIS and 89% water.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11720356 DOI: 10.1088/0031-9155/46/11/312
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609