Pace-termination and pacing for prevention of atrial tachyarrhythmias: results from a multicenter study with an implantable device for atrial therapy.

C W Israel1, B Hügl, C Unterberg, T Lawo, I Kennis, D Hettrick, S H Hohnloser.   

Abstract

INTRODUCTION: Patients with bradycardia requiring permanent pacing frequently suffer from additional atrial tachyarrhythmias (ATs). This study evaluated the safety and efficacy of atrial antitachycardia pacing (ATP) and the performance of pacing for AT prevention implemented into a new pacemaker. METHODS AND
RESULTS: In patients with conventional indications for permanent pacing, an investigational DDDRP pacemaker (Medtronic AT500, model 7253) was implanted. The primary study objectives were to determine the safety of overall device functioning and its efficacy in terminating spontaneous AT. A secondary endpoint was to determine the reliability of AT detection. Pacemaker memory functions were used to analyze the impact of dedicated pacing algorithms on AT prevention. In 33 European and Canadian centers, 325 patients were enrolled (mean follow-up 2.3+/-1.3 months). Complication-free survival at 3 months was 88%. In 2,145 episodes stored with atrial electrograms, AT detection was confirmed in 97%. The algorithm for continuous overdrive pacing increased the percentage of atrial pacing to 97%. After ATP activation, 16,683 of 52,468 AT episodes were treated (120 patients). Of these, 8,903 episodes (53%) were terminated successfully by ATP. No proarrhythmic effect of preventive pacing or atrial ATP was observed. Preventive pacing algorithms increased the median percentage of atrial pacing from 62% to 97%. However, the number of AT/AF (atrial fibrillation) episodes (4.1 vs 4.1 per patient per day) and the time in AT/AF (13.7% vs 12.8%) was not significantly different before and after activation of preventive pacing.
CONCLUSION: DDDRP pacing with a new system for AT therapy was safe and associated with successful pace-termination of AT in 53% of episodes. Preventive pacing and atrial ATP algorithms represent two new functions that can be implemented safely into pacemaker systems for nonpharmacologic treatment of ATs in patients requiring pacemaker therapy.

Entities:  

Mesh:

Year:  2001        PMID: 11699520     DOI: 10.1046/j.1540-8167.2001.01121.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  16 in total

1.  Atrial pacing for rhythm control of atrial fibrillation.

Authors:  I C van Gelder; A C P Wiesfeld
Journal:  Neth Heart J       Date:  2008-06       Impact factor: 2.380

2.  [The role of cardiac pacing in atrial arrhythmia prevention].

Authors:  G Fröhlig
Journal:  Z Kardiol       Date:  2005

3.  [Not Available].

Authors:  J C Geller; H U Klein
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2002-03

Review 4.  Management of atrial fibrillation in bradyarrhythmias.

Authors:  Giuseppe Boriani; Luigi Padeletti
Journal:  Nat Rev Cardiol       Date:  2015-03-17       Impact factor: 32.419

Review 5.  Atrial fibrillation: profit from cardiac pacing?

Authors:  A Yang; B Lüderitz; T Lewalter
Journal:  Z Kardiol       Date:  2005-03

6.  Diagnostic tools for atrial tachyarrhythmias in implantable pacemakers: a review of technical options and pitfalls.

Authors:  W G de Voogt; N M van Hemel
Journal:  Neth Heart J       Date:  2008-06       Impact factor: 2.380

7.  The usefulness of minimal ventricular pacing and preventive AF algorithms in the treatment of PAF: the 'MinVPace' study.

Authors:  Rick A Veasey; Anita Arya; Nick Freemantle; John Silberbauer; Nikhil R Patel; Guy W Lloyd; A Neil Sulke
Journal:  J Interv Card Electrophysiol       Date:  2010-01-16       Impact factor: 1.900

Review 8.  Clinical trials of pacing for maintenance of sinus rhythm.

Authors:  Anne M Gillis
Journal:  J Interv Card Electrophysiol       Date:  2004       Impact factor: 1.900

9.  Termination of atrial fibrillation using pulsed low-energy far-field stimulation.

Authors:  Flavio H Fenton; Stefan Luther; Elizabeth M Cherry; Niels F Otani; Valentin Krinsky; Alain Pumir; Eberhard Bodenschatz; Robert F Gilmour
Journal:  Circulation       Date:  2009-07-27       Impact factor: 29.690

10.  [Atrial pacing for prevention of atrial fibrillation. Influence of septal atrial pacing, atrial overdrive and AV-delay-optimization on atrial fibrillation burden].

Authors:  M Klein; N Klein; D Pfeiffer
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.