Abstract
The objectives of this study were to assess for elderly Germans the validity of existing equations for predicting body cell mass (BCM) and to develop from single- and multifrequency bioimpedance (SFBIA, MFBIA) models new prediction equations. In a data-splitting approach, validation and cross-validation were performed in 160 healthy elderly (60- to 90-yr) subjects. BCM was determined using a tetrapolar bioimpedance analyzer (800 microA; 4 fixed frequencies: 1, 5, 50, and 100 kHz; electrodes placed to hand, wrist, ankle, and foot) and whole body (40)K counting as a reference method. New prediction equations were derived by multiple stepwise regression analysis. The Bland-Altman procedure was used for methods comparison. Relative to whole body counting, the manufacturer's equation overestimated BCM by 9% in men (P < 0.0001, paired t-test) and 4% in women (P = 0.002). Compared with the manufacturer's equation, the newly derived equations (r = 0.92, RMSE = 6-9%) improved accuracy (pure error = 13 vs. 7-8%) and reduced bias and limits of agreement. SFBIA and MFBIA equations did not differ in precision or accuracy. We conclude that the newly derived equations improved BCM estimates in the elderly compared with existing equations. There was no advantage of MFBIA over SFBIA equations.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11595657 DOI: 10.1152/ajpendo.2001.281.5.E1005
Source DB: PubMed Journal: Am J Physiol Endocrinol Metab ISSN: 0193-1849 Impact factor: 4.310