Literature DB >> 11592532

Evaluation of glaucomatous visual field loss with locally condensed grids using fundus-oriented perimetry (FOP).

U Schiefer1, A Malsam, M Flad, F Stumpp, T J Dietrich, J Paetzold, R Vonthein, M Knorr, P O Denk.   

Abstract

PURPOSE: We compared detection rates of glaucomatous visual field defects (VFDs) between a conventional rectangular stimulus grid and locally condensed test point arrangements in morphologically suspicious regions.
METHODS: Humphrey Field Analyzer model 630 (HFA I, program 30-2 with a rectangular 6 degrees x 6 degrees grid) was used as the conventional perimetric method. Individual local test-point condensation was realized by fundus-oriented perimetry (FOP) on the Tuebingen Computer Campimeter (TCC).
RESULTS: Of a total of 66 glaucoma patients, or suspected sufferers, 23 showed normal findings and 27 showed pathological findings with both methods. In 15 cases we found normal visual fields in HFA 30-2, whereas FOP revealed early glaucomatous functional damage. Only one case showed pathological HFA results, while FOP was normal. Detection rates of VFDs significantly differed between the two methods (p < 0.001; sign test).
CONCLUSIONS: FOP, using individually condensed test grids, significantly increases detection rates of glaucomatous VFDs in morphologically suspicuous areas compared with a conventional HFA 30-2 technique using equidistant rectangular (6 degrees x 6 degrees) test point arrangements.

Entities:  

Mesh:

Year:  2001        PMID: 11592532     DOI: 10.1177/112067210101102s07

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  5 in total

1.  [Conventional perimetry. Antiquated or indispensable for functional glaucoma diagnostics?].

Authors:  F Tonagel; B Voykov; U Schiefer
Journal:  Ophthalmologe       Date:  2012-04       Impact factor: 1.059

2.  Novel Technique for Quantifying Retinal Nerve Fiber Bundle Abnormality in the Temporal Raphe.

Authors:  Bright S Ashimatey; Brett J King; Victor E Malinovsky; William H Swanson
Journal:  Optom Vis Sci       Date:  2018-04       Impact factor: 1.973

3.  Customizing Perimetric Locations Based on En Face Images of Retinal Nerve Fiber Bundles With Glaucomatous Damage.

Authors:  Muhammed S Alluwimi; William H Swanson; Victor E Malinovsky; Brett J King
Journal:  Transl Vis Sci Technol       Date:  2018-03-15       Impact factor: 3.283

4.  Comparison of defect depths for sinusoidal and circular perimetric stimuli in patients with glaucoma.

Authors:  William H Swanson; Brett J King
Journal:  Ophthalmic Physiol Opt       Date:  2019-01       Impact factor: 3.117

5.  Increasing the Spatial Resolution of Visual Field Tests Without Increasing Test Duration: An Evaluation of ARREST.

Authors:  Vasanth Muthusamy; Andrew Turpin; Mark J Walland; Bao N Nguyen; Allison M McKendrick
Journal:  Transl Vis Sci Technol       Date:  2020-12-16       Impact factor: 3.283

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.