Literature DB >> 11547045

Evaluation of the role of ultrasound machines as a source of nosocomial and cross-infection.

Y M Karadenz1, D Kiliç, S Kara Altan, D Altinok, S Güney.   

Abstract

RATIONALE AND
OBJECTIVES: To evaluate the potential role of ultrasonography probes and coupling gel for cross-infection in patients undergoing ultrasonography; to describe a cost-effective method of probe handling; and to emphasize the importance of position of the probes in the boxes.
METHODS: Cultures were taken from 43 probes used in different scanning regions. Cultures were also taken from probe boxes and coupling gel.
RESULTS: After abdominal scanning, cleaning with paper seems to be effective, but even wiping with alcohol is insufficient for inguinal and axillar regions. Before scanning of these areas is started, cleaning the inguinal and axillar regions with alcohol reduces the contamination risk. Placing the probes downward in boxes also increases the risk of contamination.
CONCLUSIONS: Wiping the probes with dry paper seems to be effective only for abdominal scanning; for inguinal and axillar scanning, cleaning the region with alcohol before scanning seems more appropriate. Placing the probes downward in boxes may also be a source of contamination.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11547045     DOI: 10.1097/00004424-200109000-00009

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  6 in total

1.  Superiority of chlorhexidine 2%/alcohol 70% wipes in decontaminating ultrasound equipment.

Authors:  Bhavin Shukla; Victoria Howell; Alicia Griffiths; Anita Thoppil; Monica Liu; Joseph Carter; Peter Young
Journal:  Ultrasound       Date:  2014-04-01

2.  Bacterial contamination of ultrasound probes in different radiological institutions before and after specific hygiene training: do we have a general hygienical problem?

Authors:  Thomas Sartoretti; Elisabeth Sartoretti; Candid Bucher; Aleksis Doert; Christoph Binkert; Klaus Hergan; Matthias Meissnitzer; Johannes Froehlich; Orpheus Kolokythas; Simon Matoori; Christina Orasch; Sebastian Kos; Sabine Sartoretti-Schefer; Andreas Gutzeit
Journal:  Eur Radiol       Date:  2017-04-03       Impact factor: 5.315

3.  Advancing infection control in Australasian medical ultrasound practice.

Authors:  Susan Campbell Westerway; Jocelyne M Basseal
Journal:  Australas J Ultrasound Med       Date:  2017-02-23

4.  Does the use of ultrasound affect contamination of musculoskeletal injections sites?

Authors:  Thomas Sherman; Joseph Ferguson; William Davis; Matthew Russo; Evan Argintar
Journal:  Clin Orthop Relat Res       Date:  2014-08-28       Impact factor: 4.176

Review 5.  Clinical review: Bedside lung ultrasound in critical care practice.

Authors:  Bélaïd Bouhemad; Mao Zhang; Qin Lu; Jean-Jacques Rouby
Journal:  Crit Care       Date:  2007       Impact factor: 9.097

Review 6.  Ultrasound scan as a potential source of nosocomial and crossinfection: a literature review.

Authors:  André Hadyme Miyague; Fernando Marum Mauad; Wellington de Paula Martins; Augusto César Garcia Benedetti; Ana Elizabeth Gomes de Melo Tavares Ferreira; Francisco Mauad-Filho
Journal:  Radiol Bras       Date:  2015 Sep-Oct
  6 in total