Literature DB >> 11536401

An apparatus for MR-guided breast lesion localization and core biopsy: design and preliminary results.

E Schneider1, K W Rohling, M D Schnall, R O Giaquinto, E A Morris, D Ballon.   

Abstract

System design and initial results are presented from a new unilateral MR-guided breast lesion localization and core biopsy system. Over 150 imaging studies, an accuracy study on phantoms with 50 localization wire deployments and 33 core biopsy trials, and 19 clinical procedures are reported. The mean spatial accuracy from the lesion center for a 20-gauge (G) needle (N = 13) was within 1.2 +/- 1.4 mm (SD) and for a 14G biopsy (N = 4) 0.8 +/- 1.1 mm. For sampling using a 16G core through a 14G needle, the mean accuracy was 5.6 mm (N = 2). The needle guide geometry imposed a small, calculable targeting error. For phantom measurements using the 20G device, the mean geometry-induced error was 0.73 +/- 0.43 mm. However, this contribution was, on average, 42% of the mean measured 2.35 +/- 1.65 mm offset. The new device design provided an accurate and simple guidance method for localization or core biopsy of MR-visible breast lesions. Copyright 2001 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2001        PMID: 11536401     DOI: 10.1002/jmri.1180

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  1 in total

1.  A novel MR-guided interventional device for 3D circumferential access to breast tissue.

Authors:  Matthew Smith; Xu Zhai; Ray Harter; Gale Sisney; Mai Elezabye; Sean Fain
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

  1 in total