Literature DB >> 11530061

Glaucoma drainage tube kink after pars plana insertion.

R F Rothman1, P A Sidoti, R C Gentile, J M Liebmann, H Ishikawa, R B Rosen, R Ritch.   

Abstract

PURPOSE: To describe focal obstruction of drainage tubes by kinking at the scleral entry site after pars plana insertion.
METHODS: Case study.
RESULTS: Three eyes of three patients underwent uncomplicated placement of a Baerveldt implant into the vitreous cavity at the time of pars plana vitrectomy. Intraocular pressure remained increased after the procedure without evidence of flow. Surgical exploration and modification of the tube placement resulted in immediate intraocular pressure reduction. Compression of the tube at the scleral entry site was confirmed intraoperatively in all eyes by ultrasound biomicroscopy.
CONCLUSION: Kinking of the tube at its scleral entry site should be recognized as a possible cause of increased intraocular pressure without bleb formation after pars plana insertion of a glaucoma drainage implant.

Entities:  

Mesh:

Year:  2001        PMID: 11530061     DOI: 10.1016/s0002-9394(01)00919-9

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  3 in total

1.  Pars plana-modified Ahmed Glaucoma Valve for treatment of refractory glaucoma: a pilot study.

Authors:  Torsten Schlote; Focke Ziemssen; Karl Ulrich Bartz-Schmidt
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-08-18       Impact factor: 3.117

2.  Anterior-segment imaging for assessment of glaucoma.

Authors:  Roxana Ursea; Ronald H Silverman
Journal:  Expert Rev Ophthalmol       Date:  2010-02-01

3.  Efficacy and safety of the pars plana clip in the Ahmed valve device inserted via the pars plana in patients with refractory glaucoma.

Authors:  Manuel Diaz-Llopis; David Salom; Salvador García-Delpech; Patricia Udaondo; Jose Maria Millan; J Fernando Arevalo
Journal:  Clin Ophthalmol       Date:  2010-05-06
  3 in total