Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor.

J M Wright1, G S Mattu, T L Perry, M E Gelferc, K D Strange, A Zorn, Y Chen.   

Abstract

BACKGROUND: To test the accuracy of a new algorithm for the BPM-100, an automated oscillometric blood pressure (BP) monitor, using stored data from an independently conducted validation trial comparing the BPM-100(Beta) with a mercury sphygmomanometer.
DESIGN: Raw pulse wave and cuff pressure data were stored electronically using embedded software in the BPM-100(Beta), during the validation trial. The 391 sets of measurements were separated objectively into two subsets. A subset of 136 measurements was used to develop a new algorithm to enhance the accuracy of the device when reading higher systolic pressures. The larger subset of 255 measurements (three readings for 85 subjects) was used as test data to validate the accuracy of the new algorithm.
METHODS: Differences between the new algorithm BPM-100 and the reference (mean of two observers) were determined and expressed as the mean difference +/- SD, plus the percentage of measurements within 5, 10, and 15 mmHg.
RESULTS: The mean difference between the BPM-100 and reference systolic BP was -0.16 +/- 5.13 mmHg, with 73.7% < or = 5 mmHg, 94.9% < or = 10 mmHg and 98.8% < or = 15 mmHg. The mean difference between the BPM-100 and reference diastolic BP was -1.41 +/- 4.67 mmHg, with 78.4% < or = 5 mmHg, 92.5% < or = 10 mmHg, and 99.2% < or = 15 mmHg. These data improve upon that of the BPM-100(Beta) and pass the AAMI standard, and 'A' grade BHS protocol.
CONCLUSION: This study illustrates a new method for developing and testing a change in an algorithm for an oscillometric BP monitor utilizing collected and stored electronic data and demonstrates that the new algorithm meets the AAMI standard and BHS protocol.

Entities:  

Mesh:

Year:  2001        PMID: 11518840     DOI: 10.1097/00126097-200106000-00008

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  43 in total

1.  Why use automated office blood pressure measurements in clinical practice?

Authors:  Emmanuel A Andreadis; Epameinondas T Angelopoulos; Gerasimos D Agaliotis; Athanasios P Tsakanikas; George P Mousoulis
Journal:  High Blood Press Cardiovasc Prev       Date:  2011-09-01

2.  The effect of patient-provider communication on medication adherence in hypertensive black patients: does race concordance matter?

Authors:  Antoinette Schoenthaler; John P Allegrante; William Chaplin; Gbenga Ogedegbe
Journal:  Ann Behav Med       Date:  2012-06

3.  A community-based program for cardiovascular health awareness.

Authors:  Larry W Chambers; Janusz Kaczorowski; Lisa Dolovich; Tina Karwalajtys; Heather L Hall; Beatrice McDonough; William Hogg; Barbara Farrell; Alexandra Hendriks; Cheryl Levitt
Journal:  Can J Public Health       Date:  2005 Jul-Aug

Review 4.  Recent advances in automated blood pressure measurement.

Authors:  Martin G Myers
Journal:  Curr Hypertens Rep       Date:  2008-10       Impact factor: 5.369

Review 5.  Prevalence of high blood pressure under 2017 ACC/AHA guidelines: a systematic review and meta-analysis.

Authors:  Peisheng Xiong; Zhixi Liu; Meijuan Xiong; Feng Xie
Journal:  J Hum Hypertens       Date:  2020-12-08       Impact factor: 3.012

6.  Out-of-Clinic Sympathetic Activity Is Increased in Patients With Masked Uncontrolled Hypertension.

Authors:  Mohammed Siddiqui; Eric K Judd; Byron C Jaeger; Hemal Bhatt; Tanja Dudenbostel; Bin Zhang; Lloyd J Edwards; Suzanne Oparil; David A Calhoun
Journal:  Hypertension       Date:  2019-01       Impact factor: 10.190

7.  Agency social workers could monitor hypertension in the community.

Authors:  Richard B Francoeur
Journal:  Soc Work Health Care       Date:  2010

Review 8.  The 2010 Canadian Hypertension Education Program recommendations for the management of hypertension: part I - blood pressure measurement, diagnosis and assessment of risk.

Authors:  Robert R Quinn; Brenda R Hemmelgarn; Raj S Padwal; Martin G Myers; Lyne Cloutier; Peter Bolli; Donald W McKay; Nadia A Khan; Michael D Hill; Jeff Mahon; Daniel G Hackam; Steven Grover; Thomas Wilson; Brian Penner; Ellen Burgess; Finlay A McAlister; Maxime Lamarre-Cliche; Donna McLean; Ernesto L Schiffrin; George Honos; Karen Mann; Guy Tremblay; Alain Milot; Arun Chockalingam; Simon W Rabkin; Martin Dawes; Rhian M Touyz; Kevin D Burns; Marcel Ruzicka; Norman R C Campbell; Michel Vallée; G V Ramesh Prasad; Marcel Lebel; Sheldon W Tobe
Journal:  Can J Cardiol       Date:  2010-05       Impact factor: 5.223

9.  White-Coat Effect Is Uncommon in Patients With Refractory Hypertension.

Authors:  Mohammed Siddiqui; Eric K Judd; Suzanne Oparil; David A Calhoun
Journal:  Hypertension       Date:  2017-07-10       Impact factor: 10.190

Review 10.  Automated office blood pressure measurement in primary care.

Authors:  Martin G Myers; Janusz Kaczorowski; Martin Dawes; Marshall Godwin
Journal:  Can Fam Physician       Date:  2014-02       Impact factor: 3.275

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.