Literature DB >> 1151689

Automated constant-current coulometric assay system for ascorbic acid and sodium ascorbate.

S A Moros, C M Hamilton, J E Heveran, J J Donahue, S Oliveri-Vigh.   

Abstract

The performance of an automated constant-current coulometric system for the assay of ascorbic acid and sodium ascorbate is described. After loading, it is capable of analyzing 25 samples and printing out the titer values with no operator attention for 2.5 hr. Under optimum conditions, ascertained by evaluating various electrochemical parameters, the accuracy and precision (95% ts) were found to be +/- 0.3%.

Entities:  

Mesh:

Substances:

Year:  1975        PMID: 1151689     DOI: 10.1002/jps.2600640724

Source DB:  PubMed          Journal:  J Pharm Sci        ISSN: 0022-3549            Impact factor:   3.534


  1 in total

1.  Coulometric back titration based on all-solid-state electrodes for phenylephrine hydrochloride determination.

Authors:  Qiuyu Kong; Jinghao Wu; Meng Chen; Zuanguang Chen
Journal:  Anal Bioanal Chem       Date:  2022-04-14       Impact factor: 4.142

  1 in total