Multimodality image registration quality assurance for conformal three-dimensional treatment planning.

S Mutic1, J F Dempsey, W R Bosch, D A Low, R E Drzymala, K S Chao, S M Goddu, P D Cutler, J A Purdy.   

Abstract

PURPOSE: We present a quality assurance methodology to determine the accuracy of multimodality image registration and fusion for the purpose of conformal three-dimensional and intensity-modulated radiation therapy treatment planning. Registration and fusion accuracy between any combination of computed tomography (CT), magnetic resonance (MR), and positron emission computed tomography (PET) imaging studies can be evaluated. METHODS AND MATERIALS: A commercial anthropomorphic head phantom filled with water and containing CT, MR, and PET visible targets was modified to evaluate the accuracy of multimodality image registration and fusion software. For MR and PET imaging, the water inside the phantom was doped with CuNO(3) and 18F-fluorodeoxyglucose (18F-FDG), respectively. Targets consisting of plastic spheres and pins were distributed throughout the cranium section of the phantom. Each target sphere had a conical-shaped bore with its apex at the center of the sphere. The pins had a conical extension or indentation at the free end. The contours of the spheres, sphere centers, and pin tips were used as anatomic landmark models for image registration, which was performed using affine coordinate-transformation tools provided in a commercial multimodality image registration/fusion software package. Four sets of phantom image studies were obtained: primary CT, secondary CT with different phantom immobilization, MR, and PET study. A novel CT, MR, and PET external fiducial marking system was also tested.
RESULTS: The registration of CT/CT, CT/MR, and CT/PET images allowed correlation of anatomic landmarks to within 2 mm, verifying the accuracy of the registration software and spatial fidelity of the four multimodality image sets.
CONCLUSIONS: This straightforward phantom-based quality assurance of the image registration and fusion process can be used in a routine clinical setting or for providing a working image set for development of the image registration and fusion process and new software.

Entities:  

Mesh:

Year:  2001        PMID: 11516875     DOI: 10.1016/s0360-3016(01)01659-5

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  17 in total

Review 1.  Positron emission tomography in pediatric radiation oncology: integration in the treatment-planning process.

Authors:  Matthew J Krasin; Melissa M Hudson; Sue C Kaste
Journal:  Pediatr Radiol       Date:  2004-01-27

2.  Enhancing accuracy of magnetic resonance image fusion by defining a volume of interest.

Authors:  B M Hoelper; F Soldner; R Lachner; R Behr
Journal:  Neuroradiology       Date:  2003-09-02       Impact factor: 2.804

3.  A Novel method to generate on-board 4D MRI using prior 4D MRI and on-board kV projections from a conventional LINAC for target localization in liver SBRT.

Authors:  Wendy Harris; Chunhao Wang; Fang-Fang Yin; Jing Cai; Lei Ren
Journal:  Med Phys       Date:  2018-06-13       Impact factor: 4.071

Review 4.  MRI in medical practice and its future use in radiation oncology. Resume of XXV GOCO Congress (Montpellier) 2017.

Authors:  Xavier Druet; Estrella Acosta Sanchez; Ken Soleakhena; Anne Laprie; Jordi Sáez; Stéphanie Nougaret; Olivier Riou; Elodie Rigal; Laura Kibranian; Miguel Palacios; Ismael Membrive
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-05

5.  The influence of MRI scan position on image registration accuracy, target delineation and calculated dose in prostatic radiotherapy.

Authors:  S Hanvey; A H Sadozye; M McJury; M Glegg; J Foster
Journal:  Br J Radiol       Date:  2012-12       Impact factor: 3.039

6.  Validation of the CT-MRI image registration with a dedicated phantom.

Authors:  Sofia Spampinato; Anna Maria Gueli; Luigi Raffaele; Concetta Stancampiano; Giovanni Carlo Ettorre
Journal:  Radiol Med       Date:  2014-07-15       Impact factor: 3.469

7.  Results of a multi-institutional benchmark test for cranial CT/MR image registration.

Authors:  Kenneth Ulin; Marcia M Urie; Joel M Cherlow
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-04-08       Impact factor: 7.038

8.  An evaluation of four CT-MRI co-registration techniques for radiotherapy treatment planning of prone rectal cancer patients.

Authors:  C J Dean; J R Sykes; R A Cooper; P Hatfield; B Carey; S Swift; S E Bacon; D Thwaites; D Sebag-Montefiore; A M Morgan
Journal:  Br J Radiol       Date:  2012-01       Impact factor: 3.039

9.  Effects of respiration-averaged computed tomography on positron emission tomography/computed tomography quantification and its potential impact on gross tumor volume delineation.

Authors:  Pai-Chun Melinda Chi; Osama Mawlawi; Dershan Luo; Zhongxing Liao; Homer A Macapinlac; Tinsu Pan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-07-01       Impact factor: 7.038

10.  Acceptance test of a commercially available software for automatic image registration of computed tomography (CT), magnetic resonance imaging (MRI) and 99mTc-methoxyisobutylisonitrile (MIBI) single-photon emission computed tomography (SPECT) brain images.

Authors:  Gianfranco Loi; Marco Dominietto; Irene Manfredda; Eleonora Mones; Alessandro Carriero; Eugenio Inglese; Marco Krengli; Marco Brambilla
Journal:  J Digit Imaging       Date:  2008-09       Impact factor: 4.056

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.