| Literature DB >> 11503984 |
S N Shah1, G O Knoblich, D P Lindsey, J Kreshak, S A Yerby, L B Chou.
Abstract
Intramedullary screw fixation is a popular technique for treatment of proximal fifth metatarsal fractures. The purpose of this study was to compare the fixation rigidity of a 5.5 mm partially threaded cannulated titanium screw, with presumed superior endosteal purchase, to a similar 4.5 mm screw. Acute fifth metatarsal fractures were simulated in cadavers, stabilized with intramedullary screws, and loaded to failure in three-point bending. The initial failure loads for the metatarsals fixed with 4.5 mm and 5.5 mm screws were not significantly different (332.4 N vs. 335.2 N, respectively), nor were the ultimate failure loads (849.8 N vs. 702.2 N, respectively). Based upon our results, maximizing screw diameter does not appear to be critical for fixation rigidity and may increase the risk of intraoperative or postoperative fracture.Entities:
Mesh:
Year: 2001 PMID: 11503984 DOI: 10.1177/107110070102200709
Source DB: PubMed Journal: Foot Ankle Int ISSN: 1071-1007 Impact factor: 2.827