BACKGROUND: Memory oximeters enable diagnostic studies for sleep apnea hypopnea syndrome (SAHS) to be performed in the home. However, memory capabilities may be limited. STUDY
OBJECTIVES: To compare a pulse oximeter used at home with an 8-h memory, storing data every 12 s, and in the laboratory, with on-line recording every 2 s.
DESIGN: Prospective cohort study.
SETTING: Patients' homes and a sleep laboratory. PATIENTS: One hundred patients with suspected SAHS. MEASUREMENTS: Home oximetry and a laboratory full polysomnography. The number of >/= 4% dips in pulse oximetric saturation (SpO(2)) was calculated for each study. Daytime sleepiness was assessed by the Epworth Sleepiness Scale (ESS) score.
RESULTS: The mean dips per hour were 5.3/h (range, 0 to 53/h) for home studies and 13.4/h (range, 0 to 106/h) for laboratory studies; the relationship between home and laboratory studies was as follows: home = (0.4 x laboratory) - 0.01 +/- 11.2; r(2) = 0.64. Mean difference was 8.4/h (- 2.5 to + 77.9/h), which correlated with the mean of the measurements. At a cutoff point of 10/h, 52 studies were both negative and 13 studies were both positive. Nineteen home studies were false-negatives. Sensitivity was 0.41, and specificity was 1.0. In these 19 studies, 7 patients had an ESS score > 10 and 4 patients had an ESS score > 14. To confirm that differences were due to different sampling rates, 16 additional patients had on-line data and stored data collected simultaneously in the laboratory. Mean dips per hour were 3.2/h (range, 0.1 to 18.3/h) for the stored data and 8.34/h (0.2 to 22.8/h) for on-line data; the relationship being stored was as follows: 0.5 on-line - 1.17 +/- 2.6; r(2) = 0.69. Mean difference was 5.2/h (0.04 to 15.4 h), which correlated with the mean of the measurements.
CONCLUSION: Home studies using a memory storage pulse oximeter may underestimate the number of hypoxic dips, probably due to sampling rates. Clinically significant hypoxic SAHS may therefore be missed.

Entities:  

Mesh:

Year:  2001        PMID: 11502633     DOI: 10.1378/chest.120.2.384

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  7 in total

Review 1.  Sleep . 3: Clinical presentation and diagnosis of the obstructive sleep apnoea hypopnoea syndrome.

Authors:  D Schlosshan; M W Elliott
Journal:  Thorax       Date:  2004-04       Impact factor: 9.139

2.  Comparison between a single-channel nasal airflow device and oximetry for the diagnosis of obstructive sleep apnea.

Authors:  Lydia Makarie Rofail; Keith K H Wong; Gunnar Unger; Guy B Marks; Ronald R Grunstein
Journal:  Sleep       Date:  2010-08       Impact factor: 5.849

3.  PEDIATRIC POLYSOMNOGRAPHY.

Authors:  Suzanne E Beck; Carole L Marcus
Journal:  Sleep Med Clin       Date:  2009-09

4.  Validation of the WristOx 3100 oximeter for the diagnosis of sleep apnea/hypopnea syndrome.

Authors:  Carlos Alberto Nigro; Silvia Aimaretti; Sergio Gonzalez; Edgardo Rhodius
Journal:  Sleep Breath       Date:  2008-10-02       Impact factor: 2.816

5.  Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine.

Authors:  Nancy A Collop; W McDowell Anderson; Brian Boehlecke; David Claman; Rochelle Goldberg; Daniel J Gottlieb; David Hudgel; Michael Sateia; Richard Schwab
Journal:  J Clin Sleep Med       Date:  2007-12-15       Impact factor: 4.062

6.  Comparison of Ring Pulse Oximetry Using Reflective Photoplethysmography and PSG in the Detection of OSA in Chinese Adults: A Pilot Study.

Authors:  Rui Zhao; Jianbo Xue; Xueli Zhang; Maohuan Peng; Jing Li; Bing Zhou; Long Zhao; Thomas Penzel; Meir Kryger; Xiao Song Dong; Zhancheng Gao; Fang Han
Journal:  Nat Sci Sleep       Date:  2022-08-18

7.  Pulse oximetry for the detection of obstructive sleep apnea syndrome: can the memory capacity of oxygen saturation influence their diagnostic accuracy?

Authors:  Carlos A Nigro; Eduardo Dibur; Edgardo Rhodius
Journal:  Sleep Disord       Date:  2011-10-19
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.