Inadequacy of colonoscopy revealed by three-dimensional electromagnetic imaging.

I J Adam1, Z Ali, A J Shorthouse.   

Abstract

INTRODUCTION: It is generally accepted that clinicians performing colonoscopy should reach the cecum in at least 90 percent of examinations. However, little attention has been paid to whether the endoscopist correctly estimates the amount of colon actually seen.
METHODS: During colonoscopy, endoscopists were asked to state how far they had reached. This was compared with the amount of colon actually seen, as assessed by a novel electromagnetic imaging device that recorded a three-dimensional position of the scope within a magnetic field pervading the patient's abdomen. If electromagnetic imaging showed that the cecum had not been reached, the endoscopist was asked to use the electromagnetic imaging system to determine whether it helped advance the colonoscope further.
RESULTS: In 119 patients undergoing colonoscopy, clinical assessment of position reached was correct in only 92 (77.3 percent). When the endoscopists stated that cecal landmarks had been seen (n = 85), the scope was distal to the cecum in seven cases (8.2 percent). When cecal landmarks had not been seen (n = 34), the endoscopist's assessment of the position of the scope was accurate in only 14 (41.2 percent). The use of electromagnetic imaging in this latter group assisted passage to the cecum in 26 cases (76.5 percent).
CONCLUSION: Despite assumed visualization of the cecum, inadequate colonoscopy highlights the potential for missing significant pathology in the right colon.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11496078     DOI: 10.1007/bf02235486

Source DB:  PubMed          Journal:  Dis Colon Rectum        ISSN: 0012-3706            Impact factor:   4.585


  2 in total

1.  A preliminary prospective study of the usefulness of a magnetic endoscope locating device during colonoscopy.

Authors:  S Ambardar; T D Arnell; R L Whelan; A Nihalani; K A Forde
Journal:  Surg Endosc       Date:  2005-05-05       Impact factor: 4.584

2.  New interpretation for diagnostic yield of ileoscopy: A prospective study and a brief review.

Authors:  Mohammad Hasssan Emami; Iman Saramipoor Behbahan; Hamed Daghagh Zade; Hoshang Daneshgar
Journal:  J Res Med Sci       Date:  2009-05       Impact factor: 1.852

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.