A new tool for benchmarking cardiovascular fluoroscopes.

S Balter1.   

Abstract

This paper reports the status of a new cardiovascular fluoroscopy benchmarking phantom. A joint working group of the Society for Cardiac Angiography and Interventions (SCAI) and the National Electrical Manufacturers Association (NEMA) developed the phantom. The device has been adopted as NEMA standard XR 21-2000, 'Characteristics of and Test Procedures for a Phantom to Benchmark Cardiac Fluoroscopic and Photographic Performance' in August 2000. The test ensemble includes imaging-field geometry, spatial resolution, low-contrast iodine detectability, working thickness range, motion unsharpness, and phantom entrance dose. The phantom tests systems under conditions simulating normal clinical use for fluoroscopically guided invasive and interventional procedures. Test procedures rely on trained human observers.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11487828     DOI: 10.1093/oxfordjournals.rpd.a006464