Effect of acetaminophen on the accuracy of glucose measurements obtained with the GlucoWatch biographer.

M J Tierney1, S Garg, N R Ackerman, S J Fermi, J Kennedy, M Lopatin, R O Potts, J A Tamada.   

Abstract

BACKGROUND: Improved glycemic control significantly reduces long-term microvascular complications of diabetes mellitus associated with chronic hyperglycemia. The GlucoWatch biographer is designed to facilitate intensive diabetes management by providing automatic, frequent, and noninvasive glucose readings up to three times per hour for as long as 12 hours.
METHODS: The device extracts glucose through intact skin using reverse iontophoresis and measures the extracted glucose with an electrochemical biosensor. A clinical trial was performed to assess the effect of acetaminophen, a potential interference for traditional blood glucose meters, on the accuracy of the GlucoWatch biographer in adult subjects with diabetes (n = 18). One thousand milligram doses of acetaminophen were administered to subjects in two groups: one to achieve Cmax (peak acetominophen concentration) at the time of biographer calibration and the other to achieve Cmax during the measurement period. The biographer readings were compared to serial fingerstick blood glucose measurements.
RESULTS: Time profiles over 9 hours show close tracking of the biographer glucose results with fingerstick blood glucose measurements for all groups. The mean difference between the two measurements is between 8 and 12 mg/dL for all groups. The mean absolute value of the relative difference is less than 20%, and more than 93% of the points were in the clinically acceptable (A+B) region of the Clarke Error Grid. No statistically significant differences were found for any accuracy measurement across all groups.
CONCLUSIONS: The GlucoWatch Biographer provides frequent measurements of glucose over a 12-hour period with high accuracy. No effect of therapeutic dosage of acetaminophen on the accuracy of the glucose readings was found.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11469259     DOI: 10.1089/15209150050025140

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  4 in total

1.  GlucoWatch G2 Biographer alarm reliability during hypoglycemia in children.

Authors:  Eva Tsalikian; Craig Kollman; Nelly Mauras; Stuart Weinzimer; Bruce Buckingham; Dongyuan Xing; Roy Beck; Katrina Ruedy; William Tamborlane; Rosanna Fiallo-Scharer
Journal:  Diabetes Technol Ther       Date:  2004-10       Impact factor: 6.118

2.  Carbon nanotube composites for glucose biosensor incorporated with reverse iontophoresis function for noninvasive glucose monitoring.

Authors:  Tai-Ping Sun; Hsiu-Li Shieh; Congo Tak-Shing Ching; Yan-Dong Yao; Su-Hua Huang; Chia-Ming Liu; Wei-Hao Liu; Chung-Yuan Chen
Journal:  Int J Nanomedicine       Date:  2010-05-13

3.  Accuracy of the GlucoWatch G2 Biographer and the continuous glucose monitoring system during hypoglycemia: experience of the Diabetes Research in Children Network.

Authors: 
Journal:  Diabetes Care       Date:  2004-03       Impact factor: 19.112

4.  Clinical experience of an iontophoresis based glucose measuring system.

Authors:  Sang Youl Rhee; Suk Chon; Gwanpyo Koh; Jeong Ryung Paeng; Seungjoon Oh; Jeong-taek Woo; Sung Woon Kim; Jin-Woo Kim; Young Seol Kim
Journal:  J Korean Med Sci       Date:  2007-02       Impact factor: 2.153

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.