In-vivo testing of digital cellular telephones in patients with implantable cardioverter-defibrillators.

J A Chiladakis1, P Davlouros, G Agelopoulos, A S Manolis.   

Abstract

AIMS: To investigate the susceptibility of implantable cardioverter defibrillators to electromagnetic interference generated by digital cellular telephones, functioning in both international transmission technologies: the Global System for Mobile Communication (GSM) and the Digital Cellular System (DCS 1800). METHODS AND
RESULTS: In 36 patients with transvenous implantable cardioverter defibrillators from two manufacturers (Medtronic and Guidant/CPI), cellular telephones with different levels of minimal and maximal power output were tested in the transmitting and receiving mode. Evaluation was performed in activated implantable defibrillators during spontaneous cardiac activity and continuous VVI or DDD pacing to assess possible electromagnetic interference. In two patients, appropriateness of ventricular fibrillation detection and therapy was judged during telephone testing. There was no damage, reprogramming, inappropriate shock therapy or pacing inhibition during the tests. In seven pre-pectoral Medtronic implantable defibrillators, transient electromagnetic interference caused 19 erroneous sensing events, when the operating phone was held in close vicinity to the programmer head. These 'pseudo-oversensing' events, which did not result in logging of arrhythmia episodes in the device counter, were interpreted as an adverse interaction between the telephone and the programming device.
CONCLUSION: Digital cellular telephones do not represent a risk to Medtronic and Guidant/CPI recipients of the specific implantable defibrillator models herein tested. Copyright 2001 The European Society of Cardiology.

Entities:  

Mesh:

Year:  2001        PMID: 11465966     DOI: 10.1053/euhj.2000.2516

Source DB:  PubMed          Journal:  Eur Heart J        ISSN: 0195-668X            Impact factor:   29.983


  5 in total

1.  Effects of mobile telephones on the function of implantable cardioverter defibrillators.

Authors:  Izzet Tandogan; Bulent Ozin; Huseyin Bozbas; Sibel Turhan; Ramazan Ozdemir; Ertan Yetkin; Ergun Topal
Journal:  Ann Noninvasive Electrocardiol       Date:  2005-10       Impact factor: 1.468

2.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

3.  A novel heart/trunk simulator for the study of electromagnetic interference with active implantable devices.

Authors:  A Angeloni; V Barbaro; P Bartolini; G Calcagnini; F Censi
Journal:  Med Biol Eng Comput       Date:  2003-09       Impact factor: 3.079

4.  Interference between cellular telephones and implantable rhythm devices: a review on recent papers.

Authors:  Johnson Francis; Michael Niehaus
Journal:  Indian Pacing Electrophysiol J       Date:  2006-10-01

5.  Effect of ElectroMagnetic interference from SmartPHone on cardiac ImplaNtable electronic device (EMI-PHONE study).

Authors:  Sanatcha Apakuppakul; Nilubon Methachittiphan; Sirin Apiyasawat
Journal:  J Arrhythm       Date:  2022-07-12
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.