Literature DB >> 11440432

New technique for application of fibrin sealant: rubbing method devised to prevent cerebrospinal fluid leakage from dura mater sites repaired with expanded polytetrafluoroethylene surgical membranes.

S Nakajima1, T Fukuda, M Hasue, Y Sengoku, J Haraoka, T Uchida.   

Abstract

OBJECTIVE: The effectiveness of a strong, pressure-tight method for fibrin sealant application was experimentally and clinically evaluated, using expanded polytetrafluoroethylene (ePTFE) surgical membranes.
METHODS: Two application methods were examined in two groups. For Group 1, 1 ml each of Solutions A and B was simply sprayed over the surface (spray method). For Group 2, a small amount of Solution A was rubbed into the sutured site and then Solutions A and B were sprayed over the surface (rubbing method). In the experimental study, a moment pressure elevation test and a continuous stepwise pressure elevation test were performed for both groups. The clinical study was conducted using 350 patients undergoing craniotomies. The ePTFE surgical membrane was used for 180 patients but not the other 170 patients.
RESULTS: In the experimental study, the bursting pressures were 51.4 +/- 13.0 cm H2O for Group 1 and 129 +/- 35.0 cm H2O for Group 2 in the moment pressure elevation test and 27.0 +/- 6.7 cm H2O and 100 +/- 31.9 cm H2O, respectively, in the continuous stepwise pressure elevation test. The sealing effect was significantly greater for Group 2 than for Group 1 in both tests (P < 0.001). In the clinical study using ePTFE surgical membranes, the rubbing method significantly reduced the incidence of postoperative cerebrospinal fluid leakage (1.4%), compared with the spray method (18.3%), when external decompression was not concomitantly used (P < 0.01).
CONCLUSION: The sealing effect of fibrin sealant on cerebrospinal fluid leakage was significantly greater with the rubbing method than with the spray method. The strong sealing effect obtained with the rubbing method is thought to result from firm linkage between the ePTFE surgical membrane and the fibrin sealant and also from fibrin being formed even in needle holes and spaces in the sutured site.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11440432     DOI: 10.1097/00006123-200107000-00018

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  10 in total

1.  Experimental study on effective application of fibrin glue.

Authors:  Hajime Kin; Takayuki Nakajima; Hitoshi Okabayashi
Journal:  Gen Thorac Cardiovasc Surg       Date:  2012-03-15

2.  Maximization of the sealing effect of fibrin glue in aortic surgery.

Authors:  Masaaki Naganuma; Masatoshi Akiyama; Hiroki Takaya; Kei Sakuma; Kiichiro Kumagai; Shunsuke Kawamoto; Osamu Adachi; Yoshikatsu Saiki
Journal:  Gen Thorac Cardiovasc Surg       Date:  2019-06-08

3.  Cerebrospinal fluid (CSF) leak and pseudomeningocele formation after posterior fossa tumor resection in children: a retrospective analysis.

Authors:  P Steinbok; A Singhal; J Mills; D D Cochrane; A V Price
Journal:  Childs Nerv Syst       Date:  2006-09-19       Impact factor: 1.475

4.  New application method of fibrin glue for more effective hemostasis in cardiovascular surgery: rub-and-spray method.

Authors:  Naoki Minato; Tomoki Shimokawa; Yuji Katayama; Noriko Yamada; Masayuki Sakaguchi; Manabu Itoh
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2004-08

5.  Experimental study on the viscosity and adhesive performance of exogenous liquid fibrin glue.

Authors:  Takuro Hayashi; Mitsuhiro Hasegawa; Joji Inamasu; Kazuhide Adachi; Shinya Nagahisa; Yuichi Hirose
Journal:  Neurol Med Chir (Tokyo)       Date:  2014-10-31       Impact factor: 1.742

Review 6.  The role of barrier membranes for guided bone regeneration and restoration of large bone defects: current experimental and clinical evidence.

Authors:  Rozalia Dimitriou; George I Mataliotakis; Giorgio Maria Calori; Peter V Giannoudis
Journal:  BMC Med       Date:  2012-07-26       Impact factor: 8.775

7.  Galeal tack-up sutures to prevent subgaleal cerebrospinal fluid collection.

Authors:  Won Ho Choi; Chang Taek Moon; Young-Cho Koh; Young Il Chun; Joon Cho; Sang Woo Song
Journal:  J Korean Neurosurg Soc       Date:  2013-10-31

Review 8.  Fibrin Sealant: The Only Approved Hemostat, Sealant, and Adhesive-a Laboratory and Clinical Perspective.

Authors:  William D Spotnitz
Journal:  ISRN Surg       Date:  2014-03-04

Review 9.  Cerebrospinal Fluid Leakage after Thoracic Decompression.

Authors:  Pan-Pan Hu; Xiao-Guang Liu; Miao Yu
Journal:  Chin Med J (Engl)       Date:  2016-08-20       Impact factor: 2.628

10.  Surgical nuances and placement of subgaleal drains for supratentorial procedures-a prospective analysis of efficacy and outcome in 150 craniotomies.

Authors:  Gerrit Alexander Schubert; Walid Albanna; Hussam Aldin Hamou; Konstantin Kotliar; Sonny Kian Tan; Christel Weiß; Blume Christian; Hans Clusmann
Journal:  Acta Neurochir (Wien)       Date:  2020-01-15       Impact factor: 2.216

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.