Literature DB >> 11439216

Commissioning of a commercially available system for intensity-modulated radiotherapy dose delivery with dynamic multileaf collimation.

M Essers1, M de Langen, M L Dirkx, B J Heijmen.   

Abstract

PURPOSE: To commission commercially available equipment for intensity-modulated radiotherapy (IMRT) using dynamic multileaf collimation (DMLC).
MATERIALS AND METHODS: First, the stability in leaf positioning and in realized IMRT profiles on a Varian 2300 C/D machine were determined as a function of time and gantry angle, and as a result of treatment interrupts. Second, dose distributions calculated with the CadPlan (Varian) treatment planning system, using leaf trajectories calculated with the leaf motion calculator (LMC) algorithm, were compared with distributions realized at the 2300 C/D unit.
RESULTS: Day-to-day and gantry angle variations in leaf positioning and dose delivery were very small (less than 0.1-0.2 mm and 2%). The effect of treatment interrupts on measured dose distributions was less than 2%. The agreement between the final dose distribution calculated by CadPlan and the measured dose was generally within 2%, or 2 mm at steep dose gradients, using a leaf transmission value of 1.8% and a leaf separation value of 2 mm in LMC. For narrow peaks, deviations of up to 6% were observed. LMC does not synchronize adjacent leaf trajectories resulting in tongue-and-groove underdosages of up to 29% for extreme cases.
CONCLUSIONS: The 2300 C/D machine is suitable for accurate and reproducible DMLC treatments. The agreement between dose predictions with LMC and CadPlan, and realized doses at this unit is clinically acceptable for most cases. However, differences between calculated and actual dose values may exist in peaked fluences or due to tongue-and-groove effects. Therefore, pretreatment dosimetric verification for each patient is recommended.

Entities:  

Mesh:

Year:  2001        PMID: 11439216     DOI: 10.1016/s0167-8140(01)00317-6

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  9 in total

1.  Evaluation of the Differences Between Measurements in Multiple Institutions and Calculation Modeled by Representative Beam Data in Prostate VMAT Plan.

Authors:  Hironao Goto; Hirokazu Mizuno; Yuichi Akino; Masaru Isono; Yoshihiro Tanaka; Norihisa Masai; Toshijiro Yamamoto; Masahiko Koizumi
Journal:  In Vivo       Date:  2020 May-Jun       Impact factor: 2.155

2.  Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels.

Authors:  Karunakaran Kaviarasu; N Arunai Nambi Raj; K Krishna Murthy; A Ananda Giri Babu; Bhaskar Laxman Durga Prasad
Journal:  J Med Phys       Date:  2015 Oct-Dec

3.  Verification of Dosimetric Commissioning Accuracy of Intensity Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Delivery using Task Group-119 Guidelines.

Authors:  Karunakaran Kaviarasu; N Arunai Nambi Raj; Misba Hamid; A Ananda Giri Babu; Lingampally Sreenivas; Kammari Krishna Murthy
Journal:  J Med Phys       Date:  2017 Oct-Dec

4.  IMRT clinical implementation: prostate and pelvic node irradiation using Helios and a 120-leaf multileaf collimator.

Authors:  C H Clark; C D Mubata; C A Meehan; A M Bidmead; J Staffurth; M E Humphreys; D P Dearnaley
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

5.  The dosimetric impact of control point spacing for sliding gap MLC fields.

Authors:  Benjamin J Zwan; Jonathan Hindmarsh; Erin Seymour; Kankean Kandasamy; Kirbie Sloan; Rajesakar David; Christopher Lee
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

6.  Quality control of VMAT synchronization using portal imaging.

Authors:  James L Bedford; Honorata Chajecka-Szczygielska; Michael D R Thomas
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

7.  Determination of dosimetric leaf gap using amorphous silicon electronic portal imaging device and its influence on intensity modulated radiotherapy dose delivery.

Authors:  S Timothy Peace Balasingh; I Rabi Raja Singh; K Mohamathu Rafic; S Ebenezer Suman Babu; B Paul Ravindran
Journal:  J Med Phys       Date:  2015 Jul-Sep

8.  Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study.

Authors:  P Fenoglietto; M Khodri; D Nguyen; F Josserand-Pietri; N Aillères
Journal:  Radiat Oncol       Date:  2016-01-14       Impact factor: 3.481

9.  Commissioning and quality assurance for intensity modulated radiotherapy with dynamic multileaf collimator: experience of the Pontificia Universidad Católica de Chile.

Authors:  Carlos Daniel Venencia; Pelayo Besa
Journal:  J Appl Clin Med Phys       Date:  2004-07-01       Impact factor: 2.102

  9 in total