The carbon dioxide absorption capacity of Amsorb is half that of soda lime.

H Higuchi1, Y Adachi, S Arimura, M Kanno, T Satoh.   

Abstract

UNLABELLED: A new CO(2) absorbent, Amsorb (A), which does not contain monovalent bases, is ideal because it does not degrade volatile anesthetics to either Compound A (from sevoflurane) or carbon monoxide (from desflurane, enflurane, or isoflurane). The CO(2) absorption capacity of A, however, has not been investigated under clinical conditions. In this study, we compared the longevity (time to exhaustion) and CO(2) absorption capacity (the volume of CO(2) absorbed before CO(2) rebreathing occurs) of A under low-flow anesthesia (1 L/min) with those of two soda lime absorbents-Medisorb (M) and Sodasorb (S)-by using a 750-mL ADU canister and a 1350-mL Aestiva 3000 canister. In the study with the ADU canister, the longevity of A was 213 +/- 71 min, significantly less than those of M (445 +/- 125; P < 0.01) and S (503 +/- 89; P < 0.001). The CO(2) absorption capacity (L/100 g absorbent) of A was 5.5 +/- 1.2, significantly less than those of M (10.7 +/- 1.7) and S (12.1 +/- 1.8; P < 0.001). In the study with the Aestiva 3000 canister, the longevity of A was 218 +/- 61 min, significantly less than those of M (538 +/- 136) and S (528 +/- 103; P < 0.001). The CO(2) absorption capacity (L/100 g absorbent) of A was 7.6 +/- 1.6, significantly less than those of M (14.4 +/- 1.8) and S (14.8 +/- 2.3; P < 0.001). These results indicate that the CO(2) absorption capacity of A is half that of M or S and that the difference in the CO(2) absorption capacity between A and M or S is almost constant, regardless of the canister design. IMPLICATIONS: The CO(2) absorption capacity of Amsorb is half that of Medisorb and Sodasorb under clinical low-flow (1 L/min) anesthesia with either a 750-mL Ohmeda ADU compact or a 1350-mL Ohmeda Aestiva 3000 canister.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11429370     DOI: 10.1097/00000539-200107000-00044

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  2 in total

1.  Detection of carbon monoxide production as a result of the interaction of five volatile anesthetics and desiccated sodalime with an electrochemical carbon monoxide sensor in an anesthetic circuit compared to gas chromatography.

Authors:  Christiaan Keijzer; Roberto S G M Perez; Jaap J de Lange
Journal:  J Clin Monit Comput       Date:  2007-06-28       Impact factor: 2.502

2.  In vitro performance of prefilled CO₂ absorbers with the Aisys®.

Authors:  Jan F A Hendrickx; Simon P A J De Ridder; Alexander Dehouwer; Rik Carette; Sofie De Cooman; Andre M De Wolf
Journal:  J Clin Monit Comput       Date:  2015-05-08       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.