The reliability and validity of three dimensional ultrasound volumetric measurements using an in vitro balloon and in vivo uterine model.

T Farrell1, J R Leslie, P F Chien, P Agustsson.   

Abstract

OBJECTIVE: To evaluate the reliability and validity of two and three dimensional ultrasound volumetric measurements using balloon and uterine models.
DESIGN: Prospetive observational study.
SETTING: Obstetric ultrasound department at a university teaching hospital.
METHOD: Two and three dimensional ultrasound volumetric measurements (with 5, 10 and 15 ultrasonic slices) were performed on 30 different sets of ultrasound images obtained from 15 water filled balloons with volumes ranging from 19 to 697mL. The measurements were performed independently by two observers who were blinded to the true volumes of the balloons. For the uterine model, only three dimensional ultrasonic volume measurements were performed independently on 16 uteri by two observers who were again unaware of the definitive uterine volumes. OUTCOME MEASURE: For the assessment of intra-and inter-rater reliability, the intraclass correlation coefficient was used. The index of concordance between the ultrasonic volumes and those obtained by the reference standard (validity) was assessed with the conventional Pearson's correlation coefficient, limits of agreement method and the intra-class correlation coefficient.
RESULTS: High levels of reliability and validity were obtained for both two and three dimensional ultrasound balloon volume measurements. For two dimensional ultrasonic volume measurements, the intra-class correlation coefficient ranged from 0.992 to 0.998 for reliability and validity whereas the Pearson's correlation coefficient for validity was 0.996. With three dimensional ultrasonic volume measurements, the intra-class correlation coefficient ranged from 0.991 to 0.999 for reliability and validity whereas the Pearson's correlation coefficient for validity was 0.999. Both two and three dimensional ultrasonic measurements tended to underestimate the true balloon volume with the largest observed mean difference obtained with three dimensional ultrasound measurements using five ultrasonic slices and the smallest value obtained with three dimensional ultrasound measurements employing 15 ultrasonic slices. The mean difference in volume measurement for two dimensional ultrasound was intermediate between these two values. However, two dimensional ultrasound volume measurement generated the largest range between the limits of agreement whereas the smallest range was obtained with three dimensional ultrasound using 10 ultrasonic slices. The intra-class correlation coefficient for reliability and validity with three dimensional ultrasonic uterine volume estimation ranged from 0.956 to 0.996 whereas the Pearson's correlation coefficient for validity ranged from 0.993 to 0.999). The use of three dimensional ultrasound also consistently under-estimated the actual uterine volumes. The larger the number of ultrasonic slices employed for three dimensional ultrasound, the smaller was the mean difference between the ultrasonic and true uterine volume measurements and the smaller the limits of agreement.
CONCLUSIONS: The reliability and validity of balloon and uterine volume measurement by three dimensional ultrasound is high. This allows further research on three dimensional ultrasound for measuring pelvic organ volumes in the prediction of pelvic pathology.

Entities:  

Mesh:

Year:  2001        PMID: 11426890     DOI: 10.1111/j.1471-0528.2001.00148.x

Source DB:  PubMed          Journal:  BJOG        ISSN: 1470-0328            Impact factor:   6.531


  9 in total

1.  The use of inversion mode and 3D manual segmentation in volume measurement of fetal fluid-filled structures: comparison with Virtual Organ Computer-aided AnaLysis (VOCAL).

Authors:  J P Kusanovic; J K Nien; L F Gonçalves; J Espinoza; W Lee; M Balasubramaniam; E Soto; O Erez; R Romero
Journal:  Ultrasound Obstet Gynecol       Date:  2008-02       Impact factor: 7.299

Review 2.  Three-dimensional ultrasound as a predictor of pregnancy in patients undergoing ART.

Authors:  Cemil Yaman; Richard Mayer
Journal:  J Turk Ger Gynecol Assoc       Date:  2012-06-01

3.  Reliability and responsiveness of musculoskeletal ultrasound in subjects with and without spinal cord injury.

Authors:  Shauna Dudley-Javoroski; Tara McMullen; Michelle R Borgwardt; Lauren M Peranich; Richard K Shields
Journal:  Ultrasound Med Biol       Date:  2010-10       Impact factor: 2.998

4.  Validation of multi-detector computed tomography as a non-invasive method for measuring ovarian volume in macaques (Macaca fascicularis).

Authors:  Jeryl C Jones; Susan E Appt; Stephen R Werre; Joshua C Tan; Jay R Kaplan
Journal:  Am J Primatol       Date:  2010-06       Impact factor: 2.371

5.  A noninvasive bladder sensory test supports a role for dysmenorrhea increasing bladder noxious mechanosensitivity.

Authors:  Frank F Tu; Aliza E Epstein; Kristen E Pozolo; Debra L Sexton; Alexandra I Melnyk; Kevin M Hellman
Journal:  Clin J Pain       Date:  2013-10       Impact factor: 3.442

6.  Comparison of cervical parameters by three-dimensional ultrasound according to parity and previous delivery mode.

Authors:  Yun Sung Jo; Dong Gyu Jang; Narinay Kim; Sa Jin Kim; Guisera Lee
Journal:  Int J Med Sci       Date:  2011-10-21       Impact factor: 3.738

7.  Clarifying the role of three-dimensional transvaginal sonography in reproductive medicine: an evidenced-based appraisal.

Authors:  Nick Raine-Fenning; Arthur C Fleischer
Journal:  J Exp Clin Assist Reprod       Date:  2005-08-11

8.  Therapeutic response assessment using 3D ultrasound for hepatic metastasis from colorectal cancer: Application of a personalized, 3D-printed tumor model using CT images.

Authors:  Ye Ra Choi; Jung Hoon Kim; Sang Joon Park; Bo Yun Hur; Joon Koo Han
Journal:  PLoS One       Date:  2017-08-10       Impact factor: 3.240

9.  Clinical efficacy of cervical length and volume for prediction of labor onset in VBAC candidates.

Authors:  Yun Sung Jo; Gui Se Ra Lee; Narinay Kim; Dong Gyu Jang; Sa Jin Kim; Young Lee
Journal:  Int J Med Sci       Date:  2012-10-17       Impact factor: 3.738

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.