Intraocular pressure measurement at the choroid surface: a feasibility study with implications for implantable microsystems.

R N Rizq1, W H Choi, D Eilers, M M Wright, B Ziaie.   

Abstract

AIMS: To demonstrate that a sensor, which is inserted through the sclera and placed in intimate contact with the choroid, can reliably detect changes in the intraocular pressure (IOP).
METHODS: A manometer was used to control the IOP of three cadaver eyes in steps of 7 mm Hg. A piezoresistive pressure sensor was used to measure the pressure at the choroid through a 2.5 mm diameter hole that was surgically removed from the sclera. Data were collected for two configurations; with the sensor: (i) rigidly attached to a miniature positioning stage, and (ii) sutured to the sclera.
RESULTS: Both configurations accurately tracked the manometer pressure from 10 mm Hg to 47 mm Hg. For the fixed sensor cases, the average difference between the pressure measured at the choroid and in the anterior chamber was 0.8 mm Hg for the three eyes. For the sutured sensor case, the average difference was 2.1 mm Hg-although a significant portion of this was attributed to an initial offset. The standard deviations at each pressure level for all of the choroid measurements were under 1.0 mm Hg.
CONCLUSIONS: Small changes in IOP can be accurately measured by a sensor in contact with the surface of the choroid, for both a fixed sensor configuration and for a sensor sutured to the sclera. These results are the first step in the realisation of a surgically implantable microsensor to monitor IOP for patients suffering from low tension and other difficult to manage forms of glaucoma.

Entities:  

Mesh:

Year:  2001        PMID: 11423464      PMCID: PMC1724025          DOI: 10.1136/bjo.85.7.868

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  3 in total

1.  Continuous measurement of intraocular pressure in rabbits by telemetry.

Authors:  J W McLaren; R F Brubaker; J S FitzSimon
Journal:  Invest Ophthalmol Vis Sci       Date:  1996-05       Impact factor: 4.799

2.  Radio telemetry of intraocular pressure in vitro.

Authors:  R L Cooper; D Beale
Journal:  Invest Ophthalmol Vis Sci       Date:  1977-02       Impact factor: 4.799

3.  Miniature passive pressure transensor for implanting in the eye.

Authors:  C C Collins
Journal:  IEEE Trans Biomed Eng       Date:  1967-04       Impact factor: 4.538

  3 in total
  6 in total

1.  Intraocular pressure sensor: where are we - where will we go?

Authors:  Peter Walter
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2002-04-20       Impact factor: 3.117

2.  An implantable microfluidic device for self-monitoring of intraocular pressure.

Authors:  Ismail E Araci; Baolong Su; Stephen R Quake; Yossi Mandel
Journal:  Nat Med       Date:  2014-08-24       Impact factor: 53.440

Review 3.  An overview of home tonometry and telemetry for intraocular pressure monitoring in humans.

Authors:  Edward Yung; Valerie Trubnik; L Jay Katz
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-05-29       Impact factor: 3.117

Review 4.  Development of clinically relevant implantable pressure sensors: perspectives and challenges.

Authors:  Ingelin Clausen; Thomas Glott
Journal:  Sensors (Basel)       Date:  2014-09-22       Impact factor: 3.576

Review 5.  Diurnal and 24-h Intraocular Pressures in Glaucoma: Monitoring Strategies and Impact on Prognosis and Treatment.

Authors:  Anastasios G Konstas; Malik Y Kahook; Makoto Araie; Andreas Katsanos; Luciano Quaranta; Luca Rossetti; Gábor Holló; Efstathios T Detorakis; Francesco Oddone; Dimitrios G Mikropoulos; Gordon N Dutton
Journal:  Adv Ther       Date:  2018-10-20       Impact factor: 3.845

Review 6.  Promising Approach in the Treatment of Glaucoma Using Nanotechnology and Nanomedicine-Based Systems.

Authors:  Fidiniaina Rina Juliana; Samuel Kesse; Kofi Oti Boakye-Yiadom; Hanitrarimalala Veroniaina; Huihui Wang; Meihao Sun
Journal:  Molecules       Date:  2019-10-22       Impact factor: 4.411

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.