| Literature DB >> 11418906 |
K D Dao1, E Venn-Watson, A Y Shin.
Abstract
A case of pseudoaneurysm of the radial artery after fixation of a distal radius using the AO/ASIF volar distal radius plate system is presented. This unusual complication resulted from inadequate fracture stabilization and plate preparation. Technical tips on plate shortening and preparation of the cut ends are emphasized.Entities:
Mesh:
Year: 2001 PMID: 11418906 DOI: 10.1053/jhsu.2001.24138
Source DB: PubMed Journal: J Hand Surg Am ISSN: 0363-5023 Impact factor: 2.230