Literature DB >> 11415803

A new method for evaluating motor control in gait under real-life environmental conditions. Part 1: The instrument.

R. Moe-Nilssen1.   

Abstract

OBJECTIVE: To devise an instrument to measure linear acceleration of the trunk during gait under real-life environmental conditions.
DESIGN: A mathematical algorithm is suggested to transform data to a horizontal-vertical coordinate system. Calibration procedures are described.
BACKGROUND: A portable, low mass device to measure linear acceleration of the trunk in a horizontal-vertical coordinate system during prolonged walking is of clinical interest and has not been described previously.
METHOD: Linear acceleration is registered by a triaxial, piezoresitant accelerometer, sampled at 512 Hz and collected by exchangeable 20 Mb memory cards of a body-mounted data-logger. Data are transformed to a horizontal-vertical coordinate system, utilising the accelerometer's capacity as an inclinometer. Calibration testing was done in a servohydraulic jig by measuring acceleration of 5 Hz, 16 mm vertical displacements, first along each of the sensing axis, then by a tilted accelerometer.
RESULTS: Precision testing of each axis and transformed data from all axes showed low variability for acceleration root mean square of sequences of 1000 samples (CV <0.001). Accuracy testing by double integration of transformed data from the tilted accelerometer demonstrated a mean absolute difference of 0.04 mm (SD = 0.03 mm) from the programmed input.
CONCLUSIONS: Data from the arbitrarily tilted accelerometer transformed to a horizontal-vertical coordinate system gave precise and accurate measurements of linear accelerations when tested in a servohydraulic jig. Procedures for applying the device in gait analysis are described in a companion article. RELEVANCE: Control of upper body and body centre of mass are important aspects of balance, but gait analysis has largely focused on the movements of the lower limbs. This device can be used to register acceleration of a reference point on the upper body during walking, and may thus give new information about balance in gait.

Year:  1998        PMID: 11415803     DOI: 10.1016/s0268-0033(98)00089-8

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  74 in total

1.  Portable, Non-Invasive Fall Risk Assessment in End Stage Renal Disease Patients on Hemodialysis.

Authors:  Thurmon E Lockhart; Adam T Barth; Xiaoyue Zhang; Rahul Songra; Emaad Abdel-Rahman; John Lach
Journal:  ACM Trans Comput Hum Interact       Date:  2010       Impact factor: 2.351

2.  Validation of a measure of smoothness of walking.

Authors:  Jennifer S Brach; David McGurl; David Wert; Jessie M Vanswearingen; Subashan Perera; Rakie Cham; Stephanie Studenski
Journal:  J Gerontol A Biol Sci Med Sci       Date:  2010-10-05       Impact factor: 6.053

3.  Validation of an accelerometer for determination of muscle belly radial displacement.

Authors:  T Zagar; D Krizaj
Journal:  Med Biol Eng Comput       Date:  2005-01       Impact factor: 2.602

4.  Measuring orientation of human body segments using miniature gyroscopes and accelerometers.

Authors:  H J Luinge; P H Veltink
Journal:  Med Biol Eng Comput       Date:  2005-03       Impact factor: 2.602

5.  Age-associated changes in head jerk while walking reveal altered dynamic stability in older people.

Authors:  Matthew A D Brodie; Hylton B Menz; Stephen R Lord
Journal:  Exp Brain Res       Date:  2013-10-05       Impact factor: 1.972

6.  Data fusion of body-worn accelerometers and heart rate to predict VO2max during submaximal running.

Authors:  Arne De Brabandere; Tim Op De Beéck; Kurt H Schütte; Wannes Meert; Benedicte Vanwanseele; Jesse Davis
Journal:  PLoS One       Date:  2018-06-29       Impact factor: 3.240

7.  Understanding the effects of pre-processing on extracted signal features from gait accelerometry signals.

Authors:  Alexandre Millecamps; Kristin A Lowry; Jennifer S Brach; Subashan Perera; Mark S Redfern; Ervin Sejdić
Journal:  Comput Biol Med       Date:  2015-04-04       Impact factor: 4.589

8.  Automated detection of near falls: algorithm development and preliminary results.

Authors:  Aner Weiss; Ilan Shimkin; Nir Giladi; Jeffrey M Hausdorff
Journal:  BMC Res Notes       Date:  2010-03-05

9.  Time-dependent changes in postural control in early Parkinson's disease: what are we missing?

Authors:  Silvia Del Din; Alan Godfrey; Shirley Coleman; Brook Galna; Sue Lord; Lynn Rochester
Journal:  Med Biol Eng Comput       Date:  2015-06-07       Impact factor: 2.602

10.  Role of visual input in the control of dynamic balance: variability and instability of gait in treadmill walking while blindfolded.

Authors:  Fabienne Reynard; Philippe Terrier
Journal:  Exp Brain Res       Date:  2014-12-23       Impact factor: 1.972