Abstract
Intraocular pressure (IOP) was measured by four different applanation tonometers in normal dogs. By MacKay-Marg tonometry in 391 dogs (772 eyes) the mean +/- SD IOP was 18.8 +/- 5.5 mmHg (range 8-52 mmHg). Using Tono-Pen XL tonometry in 421 dogs (823 eyes) the mean IOP was 19.2 +/- 5.9 mmHg, and the range was 4.42 mmHg. With MMAC-II tonometry in 80 dogs (158 eyes), the mean IOP was 15.7 +/- 2.8 mmHg with a range of 10-30 mmHg. By pneumatonograph tonometry in 135 dogs (255 eyes), the mean IOP was 22.9 +/- 6.1 mmHg and the range was 10-47 mmHg. In this study 53 breeds were represented. Of those breeds with six animals or more, no significant differences were detected in IOP between breeds (P > 0.353) or sex (P > 0.270). There was a significant decline of 2-4 mmHg (P > 0.0001) in IOP as age increased from less than 2 years to greater than 6 years of age. This trend was present with all of the four tonometers. There were no significant differences between the MacKay-Marg and TonoPen-XL tonometers (P > 0.198), but significant differences with the MMAC-II (P > 0.001) and pneumatonograph (P > 0.001) tonometers existed compared to the first two instruments. Based on this study and the literature, the mean IOP for the normal dog is 19.0 mmHg with a range of 11 (5%) and 29 (95%) mmHg.Entities:
Year: 1998 PMID: 11397218 DOI: 10.1046/j.1463-5224.1998.00024.x