Literature DB >> 11385376

Twenty years' experience with the Medtronic Hall valve.

E G Butchart1, H H Li, N Payne, K Buchan, G L Grunkemeier.   

Abstract

OBJECTIVE: To assess the performance of the Medtronic Hall valve (Medtronic, Inc, Minneapolis, Minn) in one institution over a 20-year period.
METHODS: Since 1979, Medtronic Hall valves have been used in 1766 procedures (736 aortic, 796 mitral, and 234 double). Patients were followed up prospectively at 6- to 12-month intervals for a total of 12,688 follow-up years. Anticoagulation data (international normalized ratio) were recorded for all patients (approximately 95,000 observations).
RESULTS: Linearized rates of valve-related late death for aortic, mitral, and double valve replacement were 0.8%/y, 0.9%/y, and 1.1%/y, respectively. Risk factors for late mortality were (relative risk) diabetes (1.9), decade of age (1.6), concomitant coronary artery bypass grafting (1.4), hypertension (1.3), non-sinus rhythm (1.3), large valve size (1.1), valve regurgitation (1.3), and male sex (1.2). For aortic, mitral, and double valve replacement, linearized rates (percent per year) of adverse events were valve thrombosis 0.04, 0.03, and 0.0; all thromboembolism 2.3, 4.0, and 3.4; stroke 0.6, 0.8, and 0.6; major hemorrhage 1.2, 1.4, and 1.6; and prosthetic endocarditis 0.4, 0.4, and 0.7. Risk factors for thromboembolism were (relative risk) mitral valve replacement (1.9), diabetes (1.8), hypertension (1.5), and history of embolism (1.4).
CONCLUSION: At 20 years the Medtronic Hall valve demonstrates excellent durability, good hemodynamic performance, and very low thrombogenicity, with a valve thrombosis rate lower than those reported for bileaflet designs. With this prosthesis, both survival and thromboembolic events are predominantly determined by patient risk factors.

Entities:  

Mesh:

Year:  2001        PMID: 11385376     DOI: 10.1067/mtc.2001.113754

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  12 in total

Review 1.  Antithrombotic therapy for the CardioWest temporary total artificial heart.

Authors:  Christopher R Ensor; William D Cahoon; Michael A Crouch; Gundars J Katlaps; Michael L Hess; Richard H Cooke; Kyle J Gunnerson; Vigneshwar Kasirajan
Journal:  Tex Heart Inst J       Date:  2010

2.  Mechanism for cavitation of monoleaflet and bileaflet valves in an artificial heart.

Authors:  Hwansung Lee; Eisuke Tatsumi; Akihiko Homma; Tomonori Tsukiya; Yoshiyuki Taenaka
Journal:  J Artif Organs       Date:  2006       Impact factor: 1.731

3.  Evaluation of exercise-induced cerebrovascular accidents after aortic valve replacement.

Authors:  Janice L Hanson; John R Broussard; Steven J Durning; Thomas J DeGraba; Mark C Haigney; Nicholas J Fortuin; Marlene S Williams
Journal:  Mayo Clin Proc       Date:  2009-06       Impact factor: 7.616

Review 4.  Asymptomatic valvular disease: who benefits from surgery?

Authors:  Naomi F Botkin; Paula S Seth; Gerard P Aurigemma
Journal:  Curr Cardiol Rep       Date:  2005-03       Impact factor: 2.931

5.  Observation of cavitation pits on a mechanical heart valve surface in an artificial heart used in in vivo testing.

Authors:  Hwansung Lee; Akihiko Homma; Eisuke Tatsumi; Yoshiyuki Taenaka
Journal:  J Artif Organs       Date:  2009-06-18       Impact factor: 1.731

6.  Pathogenesis of paravalvular leakage as a complication occurring in the late phase after surgery.

Authors:  Fumiaki Kuwabara; Akihiko Usui; Yoshimori Araki; Yuji Narita; Shinichi Mizutani; Hideki Oshima; Yuichi Ueda
Journal:  J Artif Organs       Date:  2011-04-20       Impact factor: 1.731

7.  Device thrombogenicity emulation: a novel methodology for optimizing the thromboresistance of cardiovascular devices.

Authors:  Danny Bluestein; Gaurav Girdhar; Shmuel Einav; Marvin J Slepian
Journal:  J Biomech       Date:  2012-12-06       Impact factor: 2.712

Review 8.  Heart valve replacement: which valve for which patient?

Authors:  Joseph Huh; Faisal Bakaeen
Journal:  Curr Cardiol Rep       Date:  2006-03       Impact factor: 2.931

9.  Estimation of mechanical heart valve cavitation in a pneumatic ventricular assist device.

Authors:  Hwansung Lee; Eiki Akagawa; Akihiko Homma; Tomonori Tsukiya; Eisuke Tatsumi; Yoshiyuki Taenaka
Journal:  J Artif Organs       Date:  2007-09-20       Impact factor: 1.731

10.  Stentless aortic valve replacement in the young patient: long-term results.

Authors:  Torsten Christ; Herko Grubitzsch; Benjamin Claus; Wolfgang Konertz
Journal:  J Cardiothorac Surg       Date:  2013-04-08       Impact factor: 1.637

View more