Literature DB >> 11384569

Role of frequency doubling perimetry in detecting neuro-ophthalmic visual field defects.

D Thomas1, R Thomas, J P Muliyil, R George.   

Abstract

PURPOSE: To report the ability of frequency doubling perimetry to detect "neuro-ophthalmic" field defects, characterize them as hemianopic or quadrantanopic, and differentiate glaucomatous from "other" neuro-ophthalmic field defects.
METHODS: Sixty eyes of 30 normal subjects, 50 eyes of 29 patients with glaucomatous defects, and 138 eyes of 103 patients with "typical" neuro-ophthalmic field defects underwent automated perimetry using the Swedish Interactive Threshold Algorithm and frequency doubling perimetry. The sensitivity and specificity for identification of a field defect (frequency doubling perimetry 20-5 and 20-1 screening tests), or to characterize hemianopia/quadrantanopia (full threshold test) were determined. Ability to discriminate glaucomatous defects was determined by comparing frequency doubling perimetry full threshold test in glaucoma to pooled results of normal and neuro-ophthalmic groups.
RESULTS: On frequency doubling perimetry, a single point depressed to less than 1% probability had a sensitivity of 97.1% (20-5 test) and 95.7% (20-1 test) for detecting a neuro-ophthalmic visual field defect. The corresponding specificities were 95% using pooled results in normal subjects and patients with glaucoma and "other" neuro-ophthalmic field defects. In 20-5 screening a single abnormal point depressed to less than 2% probability level had a sensitivity of 98.6% (specificity 85%). Two abnormal points in the 20-1 screening depressed to less than 1% probability level had a specificity of 100% (sensitivity 84.8%). In frequency doubling perimetry full threshold, sensitivity and specificity for detection of hemianopia were 86.8% and 83.2%; for quadrantanopia they were 79.2% and 38.6%. The sensitivity and specificity for categorizing a defect as glaucomatous were 86% and 74.7%.
CONCLUSIONS: Frequency doubling perimetry is a sensitive and specific test for detecting "neuro-ophthalmic" field defects. The presence of two abnormal points (20-1 screening program) "rules in" the presence of a field defect. A normal 20-5 program (absence of a single abnormal point) almost "rules out" a defect. Frequency doubling perimetry could not accurately categorize hemianopic, quadrantanopic, or glaucomatous defects.

Entities:  

Mesh:

Year:  2001        PMID: 11384569     DOI: 10.1016/s0002-9394(01)00837-6

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  14 in total

1.  Frequency doubling technology perimetry for neuro-ophthalmological diseases.

Authors:  C A Johnson
Journal:  Br J Ophthalmol       Date:  2004-10       Impact factor: 4.638

2.  Prevalence of crowded optic discs in adult Chinese. The Beijing Eye Study.

Authors:  Qi Sheng You; Liang Xu; Jost B Jonas
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-04-04       Impact factor: 3.117

3.  Meta-analysis of the diagnostic efficiency of frequency-doubling technology for primary glaucoma.

Authors:  Ting Liu; Xiangge He
Journal:  Front Med China       Date:  2007-02

4.  Screening for Patients with Mild Alzheimer Disease Using Frequency Doubling Technology Perimetry.

Authors:  Umit Aykan; M Orcun Akdemir; Ozlem Yildirim; Figen Varlibas
Journal:  Neuroophthalmology       Date:  2013-11-19

5.  Using frequency-doubling perimetry to detect optic neuropathy in patients with Graves' orbitopathy.

Authors:  Mário L R Monteiro; André L F Portes; Frederico C Moura; Dina B W Regensteiner
Journal:  Jpn J Ophthalmol       Date:  2008-12-17       Impact factor: 2.447

6.  Frequency doubling technology perimetry in non-arteritic ischaemic optic neuropathy with altitudinal defects.

Authors:  C A Girkin; G McGwin; J DeLeon-Ortega
Journal:  Br J Ophthalmol       Date:  2004-10       Impact factor: 4.638

Review 7.  Visual fields in neuro-ophthalmology.

Authors:  Sachin Kedar; Deepta Ghate; James J Corbett
Journal:  Indian J Ophthalmol       Date:  2011 Mar-Apr       Impact factor: 1.848

8.  Comparison of Humphrey Matrix frequency doubling technology to standard automated perimetry in neuro-ophthalmic disease.

Authors:  Michael K Yoon; Thomas N Hwang; Shelley Day; Jenny Hong; Travis Porco; Timothy J McCulley
Journal:  Middle East Afr J Ophthalmol       Date:  2012 Apr-Jun

9.  Five year incidence of visual field loss in adult Chinese. The Beijing Eye Study.

Authors:  Ya Xing Wang; Liang Xu; Xiu Ying Sun; Yang Zou; Hai Tao Zhang; Jost B Jonas
Journal:  PLoS One       Date:  2012-05-18       Impact factor: 3.240

10.  [Not Available].

Authors:  Carlo Aleci; Tiziana Usai
Journal:  Open Ophthalmol J       Date:  2008-11-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.