End tidal carbon dioxide monitoring in spontaneously breathing, nonintubated patients. A clinical comparison between conventional sidestream and microstream capnometers.

A Casati1, G Gallioli, R Passaretta, M Scandroglio, E Bignami, G Torri.   

Abstract

BACKGROUND: To evaluate the end tidal carbon dioxide estimation in nonintubated, spontaneously breathing patients using either conventional sidestream or microstream capnometers.
METHODS: Patients received a regional anesthesia technique, while the end tidal carbon dioxide partial pressure (EtCO2) was sampled through a nasal cannula (Nasal FilterLine, Nellcor, Plesanton, CA, USA) and measured using either a conventional sidestream capnometer with a 200 ml.min-1 aspiration flow rate, or a microstream capnometer (NBP-75, Nellcor Puritan Bennett, Plesanton, CA, USA) with an aspiration flow rate of 30 ml.min-1. After a 20 min period with stable hemodynamic variables (systolic arterial blood pressure within +/- 20% from baseline values), the EtCO2 was randomly recorded using one of the two capnometer while arterial blood was simultaneously drawn from the radial artery and analyzed for measurement of arterial CO2 partial pressure. Afterwards the nasal cannula was connected to the other capnometer and the procedure repeated. Both the capnometer and arterial blood gas analyzer were calibrated before each studied patient according to the manufacturer instructions. The same procedure was repeated at least two times in each patient.
RESULTS: A total of 120 pairs of EtCO2 and PaCO2 measurements were drawn from 30 adults (age: 69 +/- 5 years; weight: 70 +/- 10 kg; height: 160 +/- 10 cm): 60 using the conventional sidestream capnometer and 60 with the microstream one. The median arterial to end tidal CO2 tension difference was 4.4 mmHg (range: 0.28 mmHg) with the microstream capnometer and 7 mm Hg (range: 0-22 mmHg) with the conventional capnometer (p = 0.02).
CONCLUSION: The microstream capnometer provides a more accurate end tidal CO2 partial pressure measurement in nonintubated, spontaneously breathing patients than conventional sidestream capnometers, allowing for adequate monitoring of the respiratory function in nonintubated patients.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11376503

Source DB:  PubMed          Journal:  Minerva Anestesiol        ISSN: 0375-9393            Impact factor:   3.051


  6 in total

Review 1.  Analgesia and sedation for painful interventions in children and adolescents.

Authors:  Christoph Neuhäuser; Bendicht Wagner; Matthias Heckmann; Markus A Weigand; Klaus-Peter Zimmer
Journal:  Dtsch Arztebl Int       Date:  2010-04-09       Impact factor: 5.594

2.  A flow-through capnometer for obstructive sleep apnea.

Authors:  Shinji Yamamori; Yuji Takasaki; Makoto Ozaki; Hiroshi Iseki
Journal:  J Clin Monit Comput       Date:  2008-05-28       Impact factor: 2.502

3.  Capnography and the bispectral index-their role in pediatric sedation: a brief review.

Authors:  Maria Sammartino; Barbara Volpe; Fabio Sbaraglia; Rossella Garra; Alessandro D'Addessi
Journal:  Int J Pediatr       Date:  2010-10-03

4.  Internal carotid artery blood flow in healthy awake subjects is reduced by simulated hypovolemia and noninvasive mechanical ventilation.

Authors:  Maria Skytioti; Signe Søvik; Maja Elstad
Journal:  Physiol Rep       Date:  2016-10

5.  Evaluation of end-tidal carbon dioxide gradient as a predictor of volume responsiveness in spontaneously breathing healthy adults.

Authors:  María C Arango-Granados; Virginia Zarama Córdoba; Andrés M Castro Llanos; Luis A Bustamante Cristancho
Journal:  Intensive Care Med Exp       Date:  2018-07-30

6.  Comparing the novel microstream and the traditional mainstream method of end-tidal CO2 monitoring with respect to PaCO2 as gold standard in intubated critically ill children.

Authors:  Muhterem Duyu; Anıl Dogan Bektas; Zeynep Karakaya; Meral Bahar; Aybuke Gunalp; Yasemin Mocan Caglar; Meryem Nihal Yersel; Ozlem Bozkurt
Journal:  Sci Rep       Date:  2020-12-16       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.