Literature DB >> 11372051

In vitro mechanical integrity of hydroxyapatite coatings on Ti-6Al-4V implants under shear loading.

C Zhang1, Y Leng, J Chen.   

Abstract

A new test method for the mechanical behavior of coatings on metal substrates under shear loading has been developed. Finite element simulations show that this new test method provides an almost identical shear load on the coatings to that of the conventional test. Using the new method, the static and fatigue behavior of plasma-sprayed hydroxyapatite (HA) coatings on Ti-6Al-4V substrates were studied at room temperature as a function of processing conditions. The results of the static tests show that the nominal interface shear strength is in the range of 25 approximately 40 MPa. The fatigue resistance to cyclic shear loading was characterized by shear stress amplitude versus cycle-to-failure for the samples that failed within 1010(7)cycles, and by residual nominal interfacial shear strength for the samples that survived 1010(7)cycles. The experimental results indicate that a threshold level of shear stress amplitude exists for introducing fatigue damage to HA coatings. AES and XPS studies indicated that bonding between the coating and substrate does not occur over about 12 percent of the area of the interface while bonding in the contact area of the interface is mainly mechanical interlocking. A failure mechanism of interfacial microflaw coalescence is suggested. Copyright 2001 John Wiley & Sons, Inc. J Biomed Mater Res 56: 342-350, 2001

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11372051     DOI: 10.1002/1097-4636(20010905)56:3<342::aid-jbm1102>3.0.co;2-f

Source DB:  PubMed          Journal:  J Biomed Mater Res        ISSN: 0021-9304


  1 in total

1.  Torsional fatigue resistance of plasma sprayed HA coating on Ti-6Al-4V.

Authors:  Liling Yan; Yang Leng; Jiyong Chen
Journal:  J Mater Sci Mater Med       Date:  2003-04       Impact factor: 3.896

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.