A comparison of vertical force and temporal parameters produced by an in-shoe pressure measuring system and a force platform.

S Barnett1, J L Cunningham, S West.   

Abstract

OBJECTIVE: To investigate the ability of Pedar in-shoe system to measure vertical force accurately, by comparing it with the Kistler force platform.
DESIGN: In vivo experiment in normal subjects.
BACKGROUND: It has been suggested Pedar is highly reliable, but absolute accuracy of the system with regard to force measurement has not been comprehensively tested.
METHODS: Sampling at 99 Hz, using five healthy subjects, simultaneous data were collected barefoot, and in three types of shoes (Trainers, Oxfords, Slip-on Deck Type). Six variables obtained from the force/time curve from each footstep were compared.
RESULTS: The temporal data recorded by Pedar correlated well with that obtained using Kistler, with significant differences only in overall duration of the step in Deck shoes (P < 0.001) and Oxford shoes (P < 0.01), and peak to peak barefoot (P < 0.01). Pedar recorded a lower first peak force and mid-peak force in all cases (P < 0.001). However, the magnitude of the 2nd peak force recorded by both systems was significantly different only in Trainers (P < 0.05) and Oxford shoes (P < 0.001). The impulse data obtained with Oxford shoes was not significantly different, although barefoot, Trainers and Deck shoes were significantly lower (P < 0.001) for Pedar.
CONCLUSIONS: In most cases, comparison of data recorded by the two systems provided good evidence for the accuracy and reliability of temporal measurements and second peak force measurements taken with the Pedar in-shoe system. Relevance. In-shoe pressure data provides evidence for clinical decisions if the systems utilised are proven to be valid, repeatable and accurate. Comparison with an established force platform enables some assessment of these factors.

Mesh:

Year:  2001        PMID: 11358623     DOI: 10.1016/s0268-0033(01)00026-2

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  16 in total

1.  Chin force in violin playing.

Authors:  Satoshi Obata; Hiroshi Kinoshita
Journal:  Eur J Appl Physiol       Date:  2011-09-28       Impact factor: 3.078

2.  Increased medial foot loading during drop jump in subjects with patellofemoral pain.

Authors:  Michael S Rathleff; Camilla Richter; Christoffer Brushøj; Jesper Bencke; Thomas Bandholm; Per Hölmich; Kristian Thorborg
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-03-22       Impact factor: 4.342

3.  Withdrawal reflexes examined during human gait by ground reaction forces: site and gait phase dependency.

Authors:  Jonas Emborg; Erika G Spaich; Ole K Andersen
Journal:  Med Biol Eng Comput       Date:  2008-10-01       Impact factor: 2.602

4.  Is Plantar Loading Altered During Repeated Sprints on Artificial Turf in International Football Players?

Authors:  Olivier Girard; Grégoire P Millet; Athol Thomson; Franck Brocherie
Journal:  J Sports Sci Med       Date:  2018-08-14       Impact factor: 2.988

5.  Gait event detection using inertial measurement units in people with transfemoral amputation: a comparative study.

Authors:  Emeline Simonetti; Coralie Villa; Joseph Bascou; Giuseppe Vannozzi; Elena Bergamini; Hélène Pillet
Journal:  Med Biol Eng Comput       Date:  2019-12-23       Impact factor: 2.602

6.  Reliability of the TekScan MatScan(R) system for the measurement of plantar forces and pressures during barefoot level walking in healthy adults.

Authors:  Gerard V Zammit; Hylton B Menz; Shannon E Munteanu
Journal:  J Foot Ankle Res       Date:  2010-06-18       Impact factor: 2.303

7.  Validation of Instrumented Football Shoes to Measure On-Field Ground Reaction Forces.

Authors:  Alexandre Karamanoukian; Jean-Philippe Boucher; Romain Labbé; Nicolas Vignais
Journal:  Sensors (Basel)       Date:  2022-05-11       Impact factor: 3.847

8.  A wireless flexible sensorized insole for gait analysis.

Authors:  Simona Crea; Marco Donati; Stefano Marco Maria De Rossi; Calogero Maria Oddo; Nicola Vitiello
Journal:  Sensors (Basel)       Date:  2014-01-09       Impact factor: 3.576

9.  Contoured, prefabricated foot orthoses demonstrate comparable mechanical properties to contoured, customised foot orthoses: a plantar pressure study.

Authors:  Anthony C Redmond; Karl B Landorf; Anne-Maree Keenan
Journal:  J Foot Ankle Res       Date:  2009-06-16       Impact factor: 2.303

10.  Estimating youth locomotion ground reaction forces using an accelerometer-based activity monitor.

Authors:  Jennifer M Neugebauer; David A Hawkins; Laurel Beckett
Journal:  PLoS One       Date:  2012-10-25       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.