Literature DB >> 11357470

Recording of subcutaneous glucose dynamics by a viscometric affinity sensor.

U Beyer1, D Schäfer, A Thomas, H Aulich, U Haueter, B Reihl, R Ehwald.   

Abstract

AIMS/HYPOTHESIS: To provide a nonenzymatic sensor for glucose monitoring in subcutaneous tissue.
METHODS: A continuously working affinity sensor based on the glucose-dependent viscosity of a sensitive liquid containing dextran and concanavalin A has been designed by arranging a microdialysis probe, two flow-resisting capillaries and two pressure transducers in a linear flow system. It allows synchronous processing of the viscosity of the sensitive liquid at the standard glucose concentration and the glucose concentration to be measured. In preliminary human trials the sensor was implanted into the subcutaneous tissue of the forearm and its read-out was compared with capillary blood concentrations.
RESULTS: In vitro, the viscometric sensor shows a linear and long-term stable dependence on the glucose concentration without detectable drift. At the applied flow rate of the sensitive liquid (about 5 microliters/h) the technical delay is 5 to 10 min. The read-out of the implanted sensor followed the dynamics of the capillary blood glucose concentrations with a time-shift of 10 to 15 min but showed a systematic error when based on precalibration with polymer-free glucose solutions. After appropriate in vivo calibration, the read-out was in good or acceptable coincidence with capillary blood concentrations according to the error grid method and did not show any detectable reduction of sensitivity during the periods of measurement (up to 44 h). CONCLUSION/INTERPRETION: The viscometric-affinity sensor is an efficient tool for current research on glucose monitoring in the subcutaneous tissue and can potentially be further developed for routine clinical use.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11357470     DOI: 10.1007/s001250051637

Source DB:  PubMed          Journal:  Diabetologia        ISSN: 0012-186X            Impact factor:   10.122


  11 in total

1.  Blood glucose self-monitoring with a long-term subconjunctival glucose sensor.

Authors:  Achim Josef Müller; Monika Knuth; Katharina Sibylle Nikolaus; Roland Krivánek; Frank Küster; Christoph Hasslacher; Gerd Uwe Auffarth
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

2.  Development of novel glucose sensing fluids with potential application to microelectromechanical systems-based continuous glucose monitoring.

Authors:  Siqi Li; Xian Huang; Erin N Davis; Qiao Lin; Qian Wang
Journal:  J Diabetes Sci Technol       Date:  2008-11

3.  Periodic extraction of interstitial fluid from the site of subcutaneous insulin infusion for the measurement of glucose: a novel single-port technique for the treatment of type 1 diabetes patients.

Authors:  Werner Regittnig; Stefan Lindpointner; Stefan Korsatko; Dina Tutkur; Manfred Bodenlenz; Thomas R Pieber
Journal:  Diabetes Technol Ther       Date:  2012-11-05       Impact factor: 6.118

4.  Sensing glucose concentrations at GHz frequencies with a fully embedded Biomicro-electromechanical system (BioMEMS).

Authors:  M Birkholz; K-E Ehwald; T Basmer; P Kulse; C Reich; J Drews; D Genschow; U Haak; S Marschmeyer; E Matthus; K Schulz; D Wolansky; W Winkler; T Guschauski; R Ehwald
Journal:  J Appl Phys       Date:  2013-06-26       Impact factor: 2.546

5.  A MEMS differential viscometric sensor for affinity glucose detection in continuous glucose monitoring.

Authors:  Xian Huang; Siqi Li; Erin Davis; Charles Leduc; Yann Ravussin; Haogang Cai; Bing Song; Dachao Li; Domenico Accili; Rudolph Leibel; Qian Wang; Qiao Lin
Journal:  J Micromech Microeng       Date:  2013-05       Impact factor: 1.881

6.  Synthesis and development of poly(N-hydroxyethyl acrylamide)-ran-3-acrylamidophenylboronic acid polymer fluid for potential application in affinity sensing of glucose.

Authors:  Siqi Li; Erin N Davis; Xian Huang; Bing Song; Rebecca Peltzman; David M Sims; Qiao Lin; Qian Wang
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

7.  A MEMS Dielectric Affinity Glucose Biosensor.

Authors:  Xian Huang; Siqi Li; Erin Davis; Dachao Li; Qian Wang; Qiao Lin
Journal:  J Microelectromech Syst       Date:  2013-06-20       Impact factor: 2.417

8.  Glucose Measurement by Affinity Sensor and Pulsed Measurements of Fluidic Resistances: Proof of Principle.

Authors:  Uwe Beyer; Thomas Wyss; Franck Robin; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2014-01-01

9.  Development of boronic acid grafted random copolymer sensing fluid for continuous glucose monitoring.

Authors:  Siqi Li; Erin N Davis; Jordan Anderson; Qiao Lin; Qian Wang
Journal:  Biomacromolecules       Date:  2009-01-12       Impact factor: 6.988

10.  Micro-Electromechanical Affinity Sensor for the Monitoring of Glucose in Bioprocess Media.

Authors:  Lorenz Theuer; Micha Lehmann; Stefan Junne; Peter Neubauer; Mario Birkholz
Journal:  Int J Mol Sci       Date:  2017-06-08       Impact factor: 5.923

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.