Literature DB >> 11355814

Preliminary investigation of the application of on-line membrane extraction of trifluoroacetic acid as an aid to improvement of negative ion electrospray mass spectrometry data.

A P New1, J C Wolff, S Crabtree, L Freitas do Santos, G Okafo, J Lee, K Divan.   

Abstract

We have recently investigated the biodegradation of a number of acidic aromatic compounds that give excellent chromatography using trifluoroacetic acid (TFA) based HPLC methods. Unfortunately HPLC methods using TFA are not usually compatible with detection by negative ion mass spectrometry as TFA suppresses ionisation of the analyte during the electrospray process. We present a preliminary investigation of the use of an anion-exchange micro-membrane suppressor to remove TFA on-line post column with the aim of improvement of mass spectral data using an aromatic acid as an example, Thus LC-MS using a TFA based HPLC method with negative ion mass spectral detection is shown to be possible with good sensitivity.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11355814     DOI: 10.1016/s0021-9673(00)01169-9

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  2 in total

1.  Membrane-based continuous remover of trifluoroacetic acid in mobile phase for LC-ESI-MS analysis of small molecules and proteins.

Authors:  Zhigui Zhou; Jialing Zhang; Jiawei Xing; Yu Bai; Yiping Liao; Huwei Liu
Journal:  J Am Soc Mass Spectrom       Date:  2012-04-20       Impact factor: 3.109

2.  Microfluidic ion stripper for removal of trifluoroacetic acid from mobile phases used in HILIC-MS of intact proteins.

Authors:  Sam Wouters; Sebastiaan Eeltink; Rob Haselberg; Govert W Somsen; Andrea F G Gargano
Journal:  Anal Bioanal Chem       Date:  2021-05-28       Impact factor: 4.142

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.