B B Ba1, R Etienne, D Ducint, C Quentin, M C Saux.
Abstract
A direct injection high-performance liquid chromatographic method with column switching has been developed to determine moxifloxacin in Mueller-Hinton broth. A LiChrocart 4-4 pre-column filled with a LiChrospher 100 RP 18, 5 microm and a 150 x 4.6 mm I.D. column packed with a Supelcozil ABZ+ Plus were used and led to a retention time of 5.70 min. Fluorescence detection allowed one to reach a quantification limit of 0.05 microg/ml with a 100-microl sample size. The standard curves were linear from 0.05 to 3.2 microg/ml. Intra- and inter-day imprecisions within the linearity range were < or =4.76 and < or =5.75%, respectively. The mean relative errors for the same day and the day-to-day inaccuracies ranged from -2.93 to +4.50% and from -1.10 to +6.00%, respectively. The method was demonstrated to be useful for pharmacokinetic-pharmacodynamic studies of moxifloxacin in an in vitro model.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11318403 DOI: 10.1016/s0378-4347(00)00591-0
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273