Literature DB >> 11308200

Fracture-embolization of duromedics valve prosthesis and microscopic uncommon lesions.

E Tatou1, M Saleh, C Eicher, R Brenot, M David.   

Abstract

We report a sudden leaflet fracture of a Duromedics mitral valve 6 years after implantation. The patient had cardiogenic shock and complained of asthenia, orthopnea, and tachycardia. Transesophageal echocardiography showed the lack of one leaflet of the prosthesis and regurgitation. An emergency mitral replacement was successfully performed. Angiographic computed tomography scan localized the sequestrum that embolized the common iliac arteries. Examination of the deficient prosthesis showed multiple lesions and, in particular, a subsurface lesion that may be characteristic of carbon pyrolytic valves.

Entities:  

Mesh:

Year:  2001        PMID: 11308200     DOI: 10.1016/s0003-4975(00)02368-7

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

1.  Prosthetic mitral valve leaflet escape.

Authors:  Darae Kim; Sin Sang Hun; In-Jeong Cho; Chi-Young Shim; Jong-Won Ha; Namsik Chung; Hyun Chul Ju; Jang Won Sohn; Geu-Ru Hong
Journal:  J Cardiovasc Ultrasound       Date:  2013-06-26

2.  Spontaneous leaflet fracture resulting in embolization from mechanical valve prostheses.

Authors:  Gijs G J van Steenbergen; Queeny H Y Tsang; Stefan M van der Heide; Michel W A Verkroost; Wilson W L Li; Wim J Morshuis
Journal:  J Card Surg       Date:  2018-12-30       Impact factor: 1.620

  2 in total