The sensitivity of indicators of thrombosis initiation to a bileaflet prosthesis leakage stimulus.

B R Travis1, U M Marzec, J T Ellis, P Davoodi, T Momin, S R Hanson, L A Harker, A P Yoganathan.   

Abstract

BACKGROUND AND AIM OF THE STUDY: The recent clinical history and experimental studies of the Medtronic Parallel (MP) valve suggest that bileaflet valve leakage flow is a primary initiator of thrombosis. These studies investigated the effects of physiologic leakage flow through a MP valve on various markers of blood damage.
METHODS: A centrifugal pump was used to drive whole, human blood anticoagulated with PPACK through a circuit containing a MP 27 mm valve in the closed position (experimental runs) or a MP 27 mm valve in the open position (control runs). Samples were taken at set time intervals after the start of the pump. These samples were analyzed by cell counting, flow cytometry, and ELISA.
RESULTS: Cell counts remained relatively constant in both the experimental and control runs. Increases in plasma hemoglobin concentration and the percentage of glycophorin A-positive fragments in the cell population were not significant in either the experimental or the control runs. Plasma platelet factor 4 activity and the percentage of the CD41-positive population which was positive for annexin V increased significantly (p <0.05) in the experimental runs compared with the control runs.
CONCLUSION: The results indicate that bileaflet valve leakage flow causes significant platelet disruption, that erythrocytes are more resistant to disruption by leakage flow than platelets and granulocytes, and that annexin V binding to platelets and plasma platelet factor 4 activity are more sensitive markers of leakage induced blood damage than plasma hemoglobin concentration.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11297211

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  5 in total

Review 1.  Biological effects of dynamic shear stress in cardiovascular pathologies and devices.

Authors:  Gaurav Girdhar; Danny Bluestein
Journal:  Expert Rev Med Devices       Date:  2008-03       Impact factor: 3.166

2.  Design of a pulsatile flow facility to evaluate thrombogenic potential of implantable cardiac devices.

Authors:  Sivakkumar Arjunon; Pablo Hidalgo Ardana; Neelakantan Saikrishnan; Shalv Madhani; Brent Foster; Ari Glezer; Ajit P Yoganathan
Journal:  J Biomech Eng       Date:  2015-02-11       Impact factor: 2.097

3.  Platelet activation due to hemodynamic shear stresses: damage accumulation model and comparison to in vitro measurements.

Authors:  Matteo Nobili; Jawaad Sheriff; Umberto Morbiducci; Alberto Redaelli; Danny Bluestein
Journal:  ASAIO J       Date:  2008 Jan-Feb       Impact factor: 2.872

4.  A novel mathematical model of activation and sensitization of platelets subjected to dynamic stress histories.

Authors:  João S Soares; Jawaad Sheriff; Danny Bluestein
Journal:  Biomech Model Mechanobiol       Date:  2013-01-29

5.  A sharp interface Lagrangian-Eulerian method for rigid-body fluid-structure interaction.

Authors:  E M Kolahdouz; A P S Bhalla; L N Scotten; B A Craven; B E Griffith
Journal:  J Comput Phys       Date:  2021-05-18       Impact factor: 4.645

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.