S Vaezy1, X Shi, R W Martin, E Chi, P I Nelson, M R Bailey, L A Crum.
Abstract
High-intensity focused ultrasound (HIFU) and conventional B-mode ultrasound (US) imaging were synchronized to develop a system for real-time visualization of HIFU treatment. The system was tested in vivo in pig liver. The HIFU application resulted in the appearance of a hyperechoic spot at the focus that faded gradually after cessation of HIFU exposure. The duration of HIFU exposure needed for a hyperechoic spot to appear, was inversely related to the HIFU intensity. The threshold intensity required to produce a hyperechoic spot in liver in < 1 s was 970 W/cm(2), in situ. At this HIFU dose, no immediate cellular damage was observed, providing a potential for pretreatment targeting. The real-time visualization method was used in hemostasis of actively bleeding internal pelvic vessels, allowing targeting and monitoring of successful treatment. Real-time US imaging may provide a useful tool for image-guided HIFU therapy.Entities: