Literature DB >> 11292401

Attempted eyelid closure affects intraocular pressure measurement.

P D Gandhi1, R Gürses-Ozden, J M Liebmann, R Ritch.   

Abstract

PURPOSE: To evaluate the effect of attempted eyelid closure on intraocular pressure measurement.
METHODS: Normal subjects underwent intraocular pressure measurement in both eyes using Goldmann applanation tonometry and Tono-pen XL (Mentor, Inc, Norwell, Massachusetts) by the same examiner holding the eyelids open, both with and without the subject simultaneously attempting forced eyelid closure. Subjects were seated during all measurements and waited 5 minutes between measurements with each instrument; the order of measurement was randomized.
RESULTS: Thirty eyes of 15 subjects (six men, nine women) were enrolled. Mean age was 30.5 +/- 5.2 years (range, 24 to 40 years). With Goldmann applanation tonometry, intraocular pressure increased in both eyes with attempted eyelid closure by a mean of 1.5 +/- 2.0 mm Hg (P =.0002, paired t test; range, -2 to 8 mm Hg). With the Tono-pen XL, intraocular pressure also increased in both eyes with attempted eyelid closure by a mean of 1.9 +/- 2.7 mm Hg (P =.0002, paired t test; range, -2 to 9 mm Hg). Tono-pen XL mean intraocular pressure values in both eyes (14.4 +/- 2.3 mm Hg) consistently overestimated those of Goldmann applanation tonometry (13.0 +/- 2.2 mm Hg) by a mean of 1.4 +/- 2.3 mm Hg.
CONCLUSIONS: Attempted forced eyelid closure is a common and statistically significant source of error in routine outpatient measurement of intraocular pressure and could influence clinical management of glaucoma.

Entities:  

Mesh:

Year:  2001        PMID: 11292401     DOI: 10.1016/s0002-9394(00)00802-3

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  9 in total

1.  Effect of a tight necktie on intraocular pressure.

Authors:  C Teng; R Gurses-Ozden; J M Liebmann; C Tello; R Ritch
Journal:  Br J Ophthalmol       Date:  2003-08       Impact factor: 4.638

2.  Risk of glaucoma among patients with benign essential blepharospasm.

Authors:  Michael S Lee; Andrew R Harrison; Daniel S Grossman; Frank A Sloan
Journal:  Ophthalmic Plast Reconstr Surg       Date:  2010 Nov-Dec       Impact factor: 1.746

3.  An implantable microfluidic device for self-monitoring of intraocular pressure.

Authors:  Ismail E Araci; Baolong Su; Stephen R Quake; Yossi Mandel
Journal:  Nat Med       Date:  2014-08-24       Impact factor: 53.440

4.  Tono-pen measurement of intraocular pressure under topical anaesthesia in full term normal newborns.

Authors:  Sagili Chandrasekhara Reddy; Rosnita Alias
Journal:  Int J Ophthalmol       Date:  2014-02-18       Impact factor: 1.779

5.  Change in intraocular pressure during point-of-care ultrasound.

Authors:  Cameron Berg; Stephanie J Doniger; Brita Zaia; Sarah R Williams
Journal:  West J Emerg Med       Date:  2015-03-06

6.  Normal values of intraocular pressure in full-term Nigerian newborns.

Authors:  Victoria Ayodeji Olatunji; Dupe Ademola-Popoola; Feyi Grace Adepoju; Omotayo Olukemi Adesiyun
Journal:  Saudi J Ophthalmol       Date:  2016-07-19

7.  Factors associated with fluctuations in repeated measurements of intraocular pressure using the Goldmann applanation tonometer in Japanese patients with primary open-angle glaucoma.

Authors:  Kiyoshi Yaoeda; Atsushi Fukushima; Motohiro Shirakashi; Atsushi Miki; Takeo Fukuchi
Journal:  Clin Ophthalmol       Date:  2018-08-17

8.  Effect of Hemifacial Spasm on Intraocular Pressure Measurement.

Authors:  Erdogan Cicik; Rengin Yildirim; Ceyhun Arici; Funda Dikkaya; Osman Sevki Arslan
Journal:  J Ophthalmol       Date:  2018-02-06       Impact factor: 1.909

9.  Role of corneal biomechanical properties in predicting of speed of myopic progression in children wearing orthokeratology lenses or single-vision spectacles.

Authors:  Kin Wan; Sin Wan Cheung; James S Wolffsohn; Janis B Orr; Pauline Cho
Journal:  BMJ Open Ophthalmol       Date:  2018-12-27
  9 in total