Narrow stereotactic beam profile measurements using N-vinylpyrrolidone based polymer gels and magnetic resonance imaging.

E Pappas1, I Seimenis, A Angelopoulos, P Georgolopoulou, M Kamariotaki-Paparigopoulou, T Maris, L Sakelliou, P Sandilos, L Vlachos.   

Abstract

In this work, polymer gel-MRI dosimetry (using VIPAR gels), radiographic film and a PinPoint ion chamber were used for profile measurements of 6 MV x-ray stereotactic beams of 5 and 10 mm diameter. The VIPAR gel-MRI method exhibited a linear dose response up to 32 Gy. VIPAR gels were found to resolve the penumbra region quite accurately, provided that the in-plane image resolution of the related T2-map is adequate (< or = 0.53 mm). T2-map slice thickness had no significant effect on beam profile data. VIPAR measurements performed with a spatial resolution of 0.13 mm provided penumbra widths (80%-20% distance) of 1.34 and 1.70 mm for the 5 and 10 mm cones respectively. These widths were found to be significantly smaller than those obtained with the film (2.23 mm for the 5 mm cone, 2.45 mm for the 10 mm cone) and PinPoint (2.25 mm for the 5 mm cone, 2.52 mm for the 10 mm cone) methods. Regarding relative depth dose measurements, good correlation between VIPAR gel and PinPoint data was observed. In conclusion, polymer gel-MRI dosimetry can provide relatively accurate profile data for very small beams used in stereotactic radiosurgery since it can overcome, to some extent, the problems related to the finite size of conventional detectors.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11277225     DOI: 10.1088/0031-9155/46/3/313

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Investigating the Electronic Portal Imaging Device for Small Radiation Field Measurements.

Authors:  Arpita Agarwal; Nikhil Rastogi; K J Maria Das; S A Yoganathan; D Udayakumar; Shaleen Kumar
Journal:  J Med Phys       Date:  2017 Apr-Jun

Review 2.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

3.  Application of high field magnetic resonance microimaging in polymer gel dosimetry.

Authors:  Agnieszka Skorupa; Aleksandra Woźnica; Mateusz Ciszek; Michał Staniszewski; Marek Kijonka; Marek Kozicki; Bożena Woźniak; Andrzej Orlef; Andrzej Polański; Łukasz Boguszewicz; Maria Sokół
Journal:  Med Phys       Date:  2020-05-15       Impact factor: 4.071

4.  Dosimetric validation for an automatic brain metastases planning software using single-isocenter dynamic conformal arcsDosimetric validation for an automatic brain metastases planning software using single-isocenter dynamic conformal arcs.

Authors:  Haisong Liu; Jun Li; Evangelos Pappas; David Andrews; James Evans; Maria Werner-Wasik; Yan Yu; Adam Dicker; Wenyin Shi
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.