Inappropriate shock therapy in a heart failure defibrillator.

T R Betts1, S Allen, P R Roberts, J M Morgan.   

Abstract

A 63-year-old male with dilated cardiomyopathy underwent implantation of a "heart failure" defibrillator capable of biventricular pacing. He received an inappropriate shock 5 hours after the procedure. Stored electrograms revealed that during each sinus beat the ventricular channel recorded up to three separate events. These resulted from far-field atrial sensing by the coronary venous lead, appropriate right ventricular sensing, then delayed left ventricular sensing (the result of left bundle branch block). As a consequence of far-field left atrial sensing the two subsequent ventricular electrograms fell within the VF zone. Following an atrial premature beat, VF detection criteria were satisfied and shock therapy delivered. Although coronary venous lead repositioning eliminated far-field atrial sensing, double counting of the widely split right and left ventricular electrograms still occurred during sinus rhythm. Shortening the programmed AV delay resulted in constant biventricular pacing with a single electrogram.

Entities:  

Mesh:

Year:  2001        PMID: 11270706     DOI: 10.1046/j.1460-9592.2001.00238.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  6 in total

Review 1.  Cardiac resynchronization: a brief synopsis part II: implant and followup methodology.

Authors:  David M Kalinchak; Mark H Schoenfeld
Journal:  J Interv Card Electrophysiol       Date:  2003-10       Impact factor: 1.900

Review 2.  Noninvasive assessment of the biventricular pacing system.

Authors:  Jonathan S Steinberg; Parimal B Maniar; Steven L Higgins; Sherie L Whiting; David B Meyer; Sergio Dubner; Abrar H Shah; David T Huang; Leslie A Saxon
Journal:  Ann Noninvasive Electrocardiol       Date:  2004-01       Impact factor: 1.468

3.  Electrocardiographic follow-up of biventricular pacemakers.

Authors:  S Serge Barold; Bengt Herweg; Michael Giudici
Journal:  Ann Noninvasive Electrocardiol       Date:  2005-04       Impact factor: 1.468

Review 4.  Advances in devices for cardiac resynchronization in heart failure.

Authors:  Chu-Pak Lau; Serge Barold; Hung-Fat Tse; Kathy Lai-Fun Lee; Hon-Wah Chan; Katherine Fan; Elaine Chau; Cheuk-Man Yu
Journal:  J Interv Card Electrophysiol       Date:  2003-10       Impact factor: 1.900

5.  Inappropriate shock delivery and biventricular pacing cardiac defibrillators.

Authors:  Brant C Liu; Rollo P Villareal; Ramesh Hariharan; Abdi Rasekh; Ali Massumi
Journal:  Tex Heart Inst J       Date:  2003

6.  Frequent ICD shocks due to double sensing in patients with bi-ventricular implantable cardioverter defibrillators.

Authors:  Amin Al-Ahmad; Paul J Wang; Munther K Homoud; N A Mark Estes; Mark S Link
Journal:  J Interv Card Electrophysiol       Date:  2003-12       Impact factor: 1.900

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.